AEROCET 531S Manual Rev H
Page 29
7.5.
Printer Format
The printer report will be generated when the
REPORT
setting is set to
PRINTER
. The
printer format for
MASS
and
COUNTS
is based on the sample
MEASURE
setting. AT and
RH will be displayed when the AT/RH probe is attached.
7.5.1. Counts Format
2013-01-22 11:40:51
Location 001
0.3u 4,620,275 M3
0.5u 3,560,226 M3
1.0u 888,731 M3
5.0u 15,542 M3
10 u 7,417 M3
AT 20 C
RH 26 %
7.5.2. Mass Format
2013-01-22 11:40:51
Location 001
PM1 1.1 µg/m3
PM2.5 3.3 µg/m3
PM4 4.6 µg/m3
PM7 5.5 µg/m3
PM10 6.7 µg/m3
TSP 10.6 µg/m3
AT 20 C
RH 26 %
8.
Maintenance
Due to the nature of the instrument, there are no customer serviceable components in
the AEROCET 531S. The case of the AEROCET 531S should never be removed or
opened for any reason.
Opening or removing the case of the AEROCET 531S voids
the warranty and may result in exposure to laser radiation, which can cause eye
injury.
8.1.
Service Schedule
Although there are no customer serviceable components in the AEROCET 531S, there
are service items which ensure the proper operation of the instrument. Table 1 shows
the service schedule for the AEROCET 531S.
Time Period
Item
Manual Section
Weekly
Zero Count Test
Monthly
Flow Rate Test
Yearly
Annual Calibration
Table 1 Service Schedule