5.8.
Set Flow Screen
The AEROCET 531S has a factory calibrated flow rate of 0.1 CFM (2.83 LPM). Use the
following procedure to adjust the flow rate when a periodic flow rate check (Section 8.3)
indicates a flow rate error greater than +/- 5%.
SET FLOW
< > TO ADJUST
ENTER TO SAVE
███████
Figure 18
– Set Flow Screen
1. Connect a reference flow meter to the inlet fitting on the top of the unit (PN 9801).
2. Access the
SET FLOW
screen (Figure 18) by pressing
MENU
then select
SET FLOW
.
The pump will start automatically when you enter the
SET FLOW
screen and stop
when you leave the screen.
3.
Press ◄ or ► to obtain a reading of 0.1 CFM (2.83 LPM) +/- 5% on the reference
flow meter.
4. Press
ENTER
to save the setting or
ESC
to cancel the change.