5
SITA Messtechnik GmbH
3
Technical Data
Device:
SITA science line t60-1 / t60-2
Memory:
16
Mbyte, non-volatile
(several hundred measurements)
Power Supply:
(100...240)V / (47...63)Hz
bzw. 4 x Mignon (AA)
max./min. Ambient Temp.
(Storage/Operation):
(-20...80)°C / (0...50)°C
max. Power Consumption:
about 3,5 VA
Total Weight
(including suitcase):
6300 g
Material of Capillary
:
PEEK
Capillary Connector:
Luer Lock Connector
Surface Tension
Measuring Range
(10...100) mN/m (dyn/cm)
Error
max. 1% f.s.
Resolution
0,1mN/m
Repeatability Error
0,5mN/m
Temperature of Liquid
Measuring Range
(-10...130)°C
Error
max. 1% f.s.
Resolution
0,1°C
Repeatability Error
0,3K
Bubble Lifetime
Controlled Range
t60-1: (1000...60000)ms / t60-2: (30...60000)ms
System Deviation
auto-mode: max. 10% / online-mode: adjustable
Resolution
1ms
Bubble Frequency
Controlled Range
t60-1: (0,008...1)Hz / t60-2: (0,008...10)Hz
System Deviation
auto-mode: max. 10% / online-mode: adjustable
Resolution
3 digits