24
SITA Messtechnik GmbH
13.3 Device Info
Device infos and diagnosis of an error will be given under
menu - info/test - warnings...
, as far as the device is capable of detecting
the kind of error:
warnings...
:
device errors
A device error is indicated by a little flashing 'w' in the title line of the display (at the upper right corner ).
serial bus sensor:
EEPROM of the sensor cannot be activated
pressure pump:
Pump is giving too little pressure
sensor ventil low:
Air throughput too low
sensor capil. low:
Pressure in capillary too low
sensor capil. high:
Pressure in capillary too high
voltage low:
Supply voltage too low
current high:
Current consumption of device too high
battery low:
Battery low
temp. device low:
Device temperature below 0 °C
temp. device high:
Device temperature above 50 °C
temp. sensor low:
Temperature sensor head below 0 °C
temp. sensor high:
Temperature sensor head above 50 °C
temp. medium low:
Temperature of sample below -100 °C
temp. medium high:
Temperature of sample over 150 °C
trigger lost:
No trigger signal (due to shock or dirty / leaky capillary)
no warnings - device ok: +++
serial number...
:
Serial number of the sensor head,
main device and software version
Device health...
:
Displays important internal data
(temperature, voltages and currents)
manufacturer
:
support adress
Symbols that indicate a low battery or temperatures which are too high or too
low will also be shown in the status line of the display alternating with the
measuring mode. That results in a flashing status line.