Displayed label
Cause
Default range
corrective measure
BT MIN ER
Block
temper-
ature
below
specified range
SetTemp-0.5
◦
C
The device still warms up or
a sudden change of ambi-
ent temperature, gas flow dis-
turbed the temperature control
loop. Wait for a couple of min-
utes. The device is used out-
side the specified ambient tem-
perature or gas temperature
range. Consider the specifica-
tions. If the error persists, send
the device to Messkonzept with
description of error.
BT MAX ER
Block
temper-
ature
above
specified range
0.5
◦
C
The device still warms up or
a sudden change of ambi-
ent temperature, gas flow dis-
turbed the temperature control
loop. Wait for a couple of min-
utes. The device is used out-
side the specified ambient tem-
perature or gas temperature
range. Consider the specifica-
tions. If the error persists, send
the device to Messkonzept with
description of error.
BU MIN ER
Bridge
voltage
below
specified
range
1V
Send
the
device
to
Messkonzept
with
descrip-
tion of error.
BU MAX ER
Bridge
voltage
above
specified
range
11V
Send
the
device
to
Messkonzept
with
descrip-
tion of error.
TC MIN ER
WLD-signal
be-
low
specified
range
500mV
Send
the
device
to
Messkonzept
with
descrip-
tion of error.
TC MAX ER
WLD-signal
above
specified
range
7000mV
Send
the
device
to
Messkonzept
with
descrip-
tion of error.
EXT. ERROR
Error
routed
from input “DIN“
(0V=no
error,
+24V=error)
Signal >14V
Check the external unit sur-
veyed by the external error.
Table 9.2: Description of System Errors