minus the hysteresis given in “R1 Hysteresis“. The hysteresis is specified in percent of
threshold.
• “R 1Mode H/L“ signalling from high to low When the measured concentration decreases
and reaches the threshold, the alarm is triggered. When the measured concentration in-
creases again above the threshold, the alarm is only reset if the value increases above
the threshold plus the hysteresis given in “R1 Hysteresis“. The hysteresis is specified in
percent of threshold
• “R 1Mode OFF“ Relay 1 is not used.
8.2.2
Relay 1 Threshold
In the “R1 Threshold“ menu the threshold value is set. The value is denotes in the currently used
measuring unit (ppm in this case). Confirm with “OK“. The threshold is independent from the
measuring range.
8.2.3
Relay 1 Hysteresis
In the “R1 Hysteresis“ menu the hysteresis value is set. The value is given in percent of the
threshold value. Confirm with “OK“.