![MessKonzept FTC400 Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/messkonzept/ftc400/ftc400_operating-manual_1775515041.webp)
Page 40 of 49
8.4.3
Reset Functions
This menu provides three reset functions and the possibility to save the current calibration pa-
rameters to the Factory Settings.
• Restart only: Restart of the software
• Factory Set.: Resets all parameters to the factory set values
• Default Set.: Resets all parameters to the default set values. CAREFUL: These values are
not suited for a properly working device.
• Save Calibration: Saves the parameters relevant to calibration. They can be retrieved by
“Factory Set.“
If a rest to Factory Settings was performing without saving the calibration parameters beforehand
a new calibration might be necessary. When the device is reset to “Default Setting “ the param-
eters relevant for the proper operation are overwritten by default values. With these values the
device will not work. The correct parameters (contact Messkonzept) have to be written to the
device again, see chapter 8.4.1 or the manual of SetApp under www.messkonzept.de
File name: Manual_FTC400_1.09KD181129CWI1V01.pdf