HDM97Pocket Quick Reference

 

 

\\

  

8

www.ibpmedical.com

The adapter is suitable for measuring pressure, temperature and conductivity in flow-
though mode. With immersion usage of the probe it is suitable for measuring temperature 
and conductivity concurrently. The preferred orientation of the adapter is vertical. The axial 
connector must point down (see picture). In flow-through mode the axial connector is to be 
used as inlet meanwhile the side connector is to be used as outlet. The connection to the 
pressure transducer inside the HDM97Pocket is accomplished by the lateral connection 
above the inlet. Before measurement it is necessary to shake the adapter slightly, so that 
air bubbles trapped inside can pass off through the outlet.

For immersion usage the probe must be unscrewed from the flow-through adapter. To 

accomplish that do not turn the upper shank, the probe may be damaged. Only turn the 

nut as shown in the picture left.

Never touch the carbon ring electrode surfaces with your fingers or other objects. Do 
not buckle the probe cable. The minimum allowed bending radius is 3 cm or 1¼ inches. 

IBPVIEW DATA ACQUISITION

For details read the user manual HDM97Pocket. Templates for the HDM97Pocket meters are 
available on the CD “HDM97Pocket”.

All brand names mentioned in this document are property of their respective owners. 

Flow-through Adapter for Conductivity  
and Temperature Measurement

For complete details on IBP products visit:

www.ibpmedical.com

IBP Medical GmbH  
Ikarusalle 15  
D 30179 Hannover  
Germany

Phone: +49 511 651647      
Fax: +49 511 652284           
eMail:  [email protected]  [email protected] 

Содержание IBP HDM97Pocket Series

Страница 1: ...M97BO HDM97BP HDM97BQ This document describes the major functions of the HDM97Pocket instrument series The complete manual can be found at http www ibpmt com medical_e meter_documents shtm and on the...

Страница 2: ...tic discharge on the connectors This can lead to substantial damage of the instrument Make sure to be completely discharged before touching the connectors or cables connected to the instrument Adjust...

Страница 3: ...USB connector is mounted on the right side of the instrument It is implemented as a Mini USB A receptacle The current consumption of the HDM97Pocket all versions can be up to 500 mA while charging the...

Страница 4: ...P BQ Target time adjustment for pressure holt test PHT TIMER All Enables disables the stopwatch function Flow BL BQ Enables disables the Flow channel pH BP enables disables the pH channel TEMP BN BO B...

Страница 5: ...and save setting or to discard this setting FLOW OK Only availble with version BQ ON OFF FLOW enabled FLOW disabled OK ESC Then to apply and save setting or to discard this setting TEMP OK Only avail...

Страница 6: ...alue Reset of both the min And max Value Return to measuring mode ESC F F Channel Select No function TEMP F Available in versions bn bo bp and bq if enabled in the basic settings Selection between uni...

Страница 7: ...x Value ESC Return to measuring display Reset accumulated volume to 0 Timer If enabled in the Basic Settings OK OK OK Afterwards Afterwards Timer Timer Timer The cycle start stop and reset can start a...

Страница 8: ...n pass off through the outlet For immersion usage the probe must be unscrewed from the flow through adapter To accomplish that do not turn the upper shank the probe may be damaged Only turn the nut as...

Отзывы: