5
III. SAFETY INFORMATION
WARNINGS AND CAUTIONS
WARNING:
The 90XL Display Module and its attached sensor modules should never be used
without first verifying accuracy of measurement functions. Using the device without verifying
accurate function could result in improper calibration of the dialysis delivery system. An
improperly calibrated dialysis delivery system could result in compromised dialysis treatment(s)
and/or patient injury.
Mesa Labs recommends verifying sensor module calibration using National Institute of
Standards and Technology (N.I.S.T.) traceable reference standards before use or whenever
inaccurate readings are suspected.
WARNING:
The device is not suitable for use in a potentially explosive environment. The
device cannot be used inside an oxygen tent.
CAUTION:
Use only sodium chloride (NaCl) standard solution. The conductivity/temperature
sensor module is temperature-compensated for NaCl solution. Other solutions, such as
potassium chloride (KCl), may result in an inaccurate calibration.
CAUTION:
DO NOT use the 90XL Instrumentation System in place of the dialysis delivery
system’s primary sensors, controls, and/or monitors.
CAUTION:
DO NOT submerse your 90XL display module or the sensor module connector
ends in liquids, as this may result in damage to the instrument and/or the sensor modules.
CAUTION:
DO NOT use abrasive cleaning agents and/or full-strength bleach or acid to clean
the display module or sensor modules, as this will cause damage.
CAUTION:
DO NOT open the display module or sensor modules. There are no user-
serviceable parts inside and it will void the warranty.
CAUTION:
Federal (USA) law restricts this device to sale to or by order of a physician.
CAUTION:
Dispose of this device properly.
It should not be disposed of in a landfill.
WARNING:
The user can be exposed to the following materials when using this device:
Polyester, Aluminum, Acrylic, Glass, Natural Polypropylene, *Noryl
TM
, Stainless Steel, Santoprene,
Silicone, ABS Plastic, Polycarbonate Plastic, Powder Coat Paint. No latex or latex containing
materials are used in the manufacture of the device.
*Noryl
TM
is a registered trademark of SABIC Innovative Plastics IP B.V.