1.
Low Range Conductivity Calibration
PROCEDURE
DISPLAY
· No reference solution is required.
· Make sure a condo/temp sensor module is attached.
After “CALIBRATION” has been entered, Select “AIR
ZERO” from the next menu list.
Step 1
– Draw DI or RO water through
the sensor module to rinse. Empty the sensor module
as completely as possible and allow to dry.
Press NEXT to continue…
NOTE:
Gently tap the cell to expel droplets, or let dry.
This initial step may be performed in advance of the
actual calibration procedure.
Step 2
– The 90XL will read the cell (theoretically
zero) and perform an evaluation based on the factory
set value. Press NEXT to continue...
Step 3
– If the evaluation is successful, the calibration
information will be stored within the sensor module.
Step 3
– If the displayed factory cal value is greater
than 0.02 mS, the AIR ZERO calibration will fail
indicating possible sensor module malfunction. Press
ABORT to exit. If the calibration fails on multiple
attempts, call Mesa Labs for assistance.