Turn the system 

OFF

 when not in use for an extended period of time. 

 

When running Postvasectomy tests do not interrupt test cycle nor interfere with system 
or testing capillary in any way – this test is highly sensitive to any motion and requires 
complete stability of the system during the 5-minute testing cycle. 

 

Variations in ambient temperature can affect semen sample parameters. 

It is essential 

that semen samples are not heated

 for testing. The SQA-Vision is calibrated to conduct 

tests at room temperature: 20-25ºC (68-77ºF). 

 

Semen  is  considered  a  biologically  hazardous  material  and  is  subject  to 
individual laboratory protocols for handling such materials and at a minimum: 

 

 

Laboratory coat, mask and gloves for operating personnel protection 

 

 

Samples handling and waste disposal in specially marked hazardous waste containers 

 

Only personnel trained to work with biologically hazardous materials such as semen 
should be testing and handling semen 

Operating Temperature and Humidity  

 

Maximum  operational  humidity  is  up  to  80%  for  temperatures  of  up  to  31

º

C  with 

decreasing linearly to 50% at 38

º

C.  

 

Operates  in  a  wide  range  of  ambient  temperatures  (15-38

º

C)  however  the  system  is 

calibrated 

to 

measure 

semen 

samples 

at 

room 

temperature:  

20-25

º

C  (68-77

º

F).    Note:    Extreme  ambient  temperature  may  impact  the  accuracy  of 

motility test results because of the known effect of temperature on human semen. 

Operational Environmental conditions: 

 

System  is  intended  for  indoor  use  at  a  maximum  altitude  of  2000m,  mains  supply 
fluctuations ±10%, Overvoltage Category II, Pollution Degree II. 

PC / Hardware  

 

PC and device:

 "All in One" - computer with SQA-Vision software and device. 

Quality Control 

 

Internal:

  Electronic  Self-Test  and  Auto-Calibration.  Runs  automatically  upon  start-up. 

Besides that, reference values are verified prior to each test.  

 

External:

  Runs QC samples daily prior to testing or per laboratory protocol. Runs assayed 

latex  bead  control: 

"QwikCheck™-beads

"  (product  of  Medical  Electronic  Systems)  for 

concentration  and  negative  control  for  motility/concentration  OR  non-assayed:  Latex 
beads or stabilized sperm CAP or NEQAS for concentration.  

  

 

Sample Testing  

 

Sample  Testing  Temperature:

  Calibrated  for  room  temperature only.  Motility  results 

will be impacted by heating the specimen (not recommended). 

 

System calibrated to test Human semen and specified Control samples only.  

Not 

for use with animal semen. 

 

SQA-Vision  measurement  capillary:

    Disposable,  plastic,  testing  capillary.  Requires 

500 µl of sample for normal volume testing, 20 µl for low volume testing, 300 µl for diluted 
mode.  Use only manufacturers’ certified testing capillaries in the automated system. 

 

Field of View Stage:

 This stage is an integral part of the SQA-Vision visualization 

compartment.  Vision™ fixed cover slip slides (obligatory for accurate test results when 
manual sperm counting is required), standard slides (for debris scans and image 
capturing) are accepted in the field of view stage using the supplied slide holder.  

 

 

Содержание SQA-Vision

Страница 1: ...V e r s i o n 1 0 9 1 3 4 C a t a l o g V S M L 0 1 0 9 1 0 0 R e v J a n 2 0 2 0...

Страница 2: ...on Start Up 9 Prior to testing a sample 10 SECTION VI Getting Started Set Up Getting Started Set Up 10 SECTION VII Troubleshooting Overview 11 SECTION VIII Technical Support and Part Replacement Instr...

Страница 3: ...nt is performed using the DEBRIS ROUND CELLS SCANNER Normal morphologically is assessed automatically along with other semen parameters Vitality and Morphology differentials can be assessed manually u...

Страница 4: ...lgorithm makes an adjustment to account for the infrared light absorption of the seminal fluid components SQA Vision Reportable Range Automated Results Sample Type Sperm Conc M ml Motility Morph MSC M...

Страница 5: ...fuses 250V 2A RS232 connector 2 USB 2 0 B Type Male On Off switch of the blue surrounding LEDs located on bottom side of the panel Left Side Panel Power On Off switch I Button port Visualization Comp...

Страница 6: ...e known effect of temperature on human semen Operational Environmental conditions System is intended for indoor use at a maximum altitude of 2000m mains supply fluctuations 10 Overvoltage Category II...

Страница 7: ...to move to the next screen or field Use the ESC button to return to the previous screen or field Rear Panel Rear Panel Operation Display I Button Keypad Automated Measurement Compartment Focus Knob F...

Страница 8: ...rtment Measurement Compartment Focus Control Field of View control Motherboard Schematic Schematic of the SQA Vision Motherboard 5 VDC 3 VDC 12 VDC 2 VDC AC 100 230 V 50 60 Hz 12 VDC 5 VDC 12 VDC 12 V...

Страница 9: ...Accessories Kit Each kit contains Power Cables Communication Cable USB cables RS232 cable RS232 to USB converter Instructional Guides Slide holder Additional set of device feet Barcode reader Capillar...

Страница 10: ...verification Reference parameters of the concentration and motility channels are measured again without a testing capillary System noise Measures the electronic noise level of the system to insure eff...

Страница 11: ...y cause damage to the calibration AND will VOID THE WARRANTY The electro optical board should NEVER be touched when the SQA Vision is opened it will cause damage to the system s calibration Technical...

Страница 12: ...lly peel off the damaged keypad from the front panel Clean the surface with alcohol and let it dry Pull the keypad cable through slit in the front panel Please note DO NOT connect it currently Attach...

Страница 13: ...ltage is evident replace the power inlet see the relevant section below If voltage is being supplied to the input connector check the output connector How to check the output voltage of PSU Disconnect...

Страница 14: ...pull the power inlet out of the SQA Vision through the rear panel Insert a new power inlet and re connect the cables Re connect the internal cables re place the internal cover and re connect the power...

Страница 15: ...re the old inverter board Replace the inverter board with a new one and secure it with the two screws Re connect the cables of the inverter board If the problem persists contact MES Customer Support S...

Страница 16: ...ns If the software was installed successfully and the problem remains check the Data flat cable Open the SQA Vision and verify that the Data flat cable with the red lined side up is connected to the s...

Страница 17: ...able which connects the main board and the Backlight PCB Item KHD 908 000858 If replacing the long flat cable does not work Replace the short flat cable which connects the LCD screen and the Backlight...

Страница 18: ...b Part VS M 01105 00 ISSUE When using the FOV knob there is no change in the field of view SOLUTIONS This is described in a separate technical bulletin please contact MES support services for further...

Страница 19: ...02 Issue When the SQA Vision is turned on it fails SELF TEST Ensure there is no testing capillary in the measurement compartment Remove the SQA Vision from sources of electronic noise cell phones etc...

Страница 20: ...ector from its location on the main board J13 Attach a new capillary sensor to the optical block using a screw Gently re attach the Dust Buster to the Optical Block using the 4 Phillips screws Re atta...

Страница 21: ...connect the optical assembly from the front panel by unscrewing 5 screws using a 2 5 Allen Key Release the screws at the rear section of the optical column using a 2 5 Allen key Release the zoom DC mo...

Страница 22: ...a assembly Slide the entire new camera assembly back into the optical assembly over the false track Re attach the Camera Stopper Reassemble the optical assembly by screwing the four screws on the rear...

Страница 23: ...SQA Vision by unscrewing the screws from the rear panel using a 2 Philips screwdriver Pull out the focus knob Using a 2 5 Allen Key release the 5 screws which hold the Optical column to the front pan...

Страница 24: ...SQA Vision Service Manual Version 109 13 4 Edition rev Jan 2020 24 SECTION IX Appendixes Appendix 1 SQA Cleaning Instructions...

Страница 25: ...t it into the correct position 4 Close the SQA Vision and tighten all screws of the rear panel 5 Connect the power cable of the SQA Vision 6 Turn the SQA Vision on and run the SELF TEST 7 If the SQA V...

Страница 26: ...e processor socket see the figure to the right 5 Slightly press the center of the processor with an index finger to re seat it into the correct position see the figure to the right 6 Close the SQA Vis...

Страница 27: ...K Problem was solved Yes Yes Yes Yes End of troubleshooting Not Not Not Not Voltage is OK Check the input voltage of PSU see p 13 Not OK Not OK Not OK Not OK Replace the power inlet see p 14 OK OK OK...

Страница 28: ...LF TEST If the LED CURRENT 1 LED 1 is not within the acceptable range of 5 20mA and is not identical to the initial manufacturer s setting you must compare it indicates the following High vs Initial S...

Страница 29: ...500mV 3 500mV C O N C E N T R A T I O N REFERENCE 2 REF 2 2500 3500 REFERENCE 2 FAILED THE SELF TEST If the REFERENCE 2 REF 2 is not within the acceptable range of 2500 3500mV it indicates the followi...

Страница 30: ...negative control is within the acceptable range as specified on the controls box Control Level 1 2 3 negative CONTROL LEVEL TEST STATUS IS NOT DEFINED If there is a probability that the control test...

Страница 31: ...ESSQAV mdf file is closed properly Call for technical assistance if the problem remains Last Backup BACKUP DESCRIPTION FILE WAS NOT SAVED THIS ERROR DOES NOT CAUSE THE BACKUP PROCESS TO FAIL SO YOU MA...

Отзывы: