SQA-Vision
Service Manual Version 109.13.4
Edition rev.: Jan 2020
21
CCD
Sensor
Camera Replacement (Part #
VS-MA-01174-00)
Issue:
When the camera is dirty or malfunctioning, then the CCD sensor including the
assembly of the mount and the dust sleeve should be replaced.
Turn off the power and open the SQA-Vision by unscrewing the screws from the rear
section of the rear panel using a #2 Philips screwdriver.
Pull out the focus knob.
Disconnect the optical assembly from the front panel by unscrewing 5 screws using a #2.5
Allen Key.
Release the screws at the rear section of the optical column
using a #2.5 Allen key
Release the zoom DC motor by unscrewing 2 screws using a
#2.5 Allen Key
Release the USB protector and the bottom cover by unscrewing 4 screws using a #2.5
Allen Key
Release the camera USB stopper by unscrewing 2 screws using #0 Phillips screwdriver
Remove the DC motor
assembly
Release motor
assembly screws
Remove the screws
at the rear section
Remove the USB protector and the bottom cover
Remove the USB stopper and disconnect
the camera USB cable