Spectroquant
®
Colorimeter Picco COD / CSB
2
Description of functions
2.2 Display
backlight
Press the [MENU] key to turn the display backlight on or off. The backlight is
switched off automatically during the measurement.
2.3 Recall of stored data
If the instrument is switched on, press the [MENU] key for more than 4
seconds to access the recall menu.
For further details see section 3.2, page 25.
2.4 Measurement of the absorbance
This mode function allows the measurement of the absorbance at the specified
wavelength.
The result appears in the display in mAbs (1000 mAbs = 1 A (Absorbance
unit) ).
Abs 1:
λ
= 430 nm
Abs 2:
λ
= 605 nm
Menu
Menu
AbS I
AbS 2