Release 03/2009
32
en
Spectroquant
®
Colorimeter Picco COD / CSB
6 Methods
6.1 Brief
instructions
•
Method notes
•
Observe the package inserts, applications, and effects of the sample matrix of the
respective test kit.
•
Ensure proper disposal of reagent solutions.
•
Material Safety Data Sheets are available on request (http://photometry.merck.de).
•
Test kits and measuring ranges
Method
Spectroquant
®
Cat. No.
Measuring
Round
Digestion
test
kit
range
cell
time
U.1
COD Cell Test
114540
10 - 150 mg/l
16 mm
120 min
U.2
COD Cell Test
114541
25 - 1500 mg/l
16 mm
120 min
U.3
COD Cell Test
114691
300 - 3500 mg/l
16 mm
120 min
U.4
COD Cell Test
114555
0.50 - 10.00 g/l
16 mm
120 min
Abs 11
Absorbance
-
-100 - 2500 mA
16 mm
-
(
λ
= 430 nm)
Abs 2
Absorbance
-
-100 - 2500 mA
16 mm
-
(
λ
= 605 nm)
Notes:
When using the 16-mm round cell the adapter for 16-mm cells must be attached to
the measurement compartment. For all measurements the measurement compart-
ment must be protected using the enclosed compartment lid.
The measurement result is displayed in mg/l (U.1 - U.3) or g/l (U.4).
Spectroquant
®
Colorimeter Picco
COD
Содержание Spectroquant Picco COD/CSB
Страница 2: ......
Страница 18: ...Ausgabedatum 03 2009 18 Spectroquant Colorimeter Picco COD CSB de...
Страница 34: ...Release 03 2009 34 en Spectroquant Colorimeter Picco COD CSB...
Страница 50: ...Version du 03 2009 50 Spectroquant Colorim tre Picco COD CSB fr...
Страница 66: ...Edizione 03 2009 66 Spectroquant Colorimetro Picco COD CSB it...
Страница 82: ...Fecha de emisi n 03 2009 82 Spectroquant Color metro Picco COD CSB sp...