Installation Note
The ATEX mixer is designed for operation in the classification noted on the product label.
The mixer must be in good working order before operating.
Before using the mixer, ensure the following:
•
The tank plate is installed properly on the tank and concentricity and alignment have been
verified to prevent any contact between the mixing head, outer rotor and tank plate. If the
tank plate is warped during welding or not aligned properly with the tank surface, remove and
discard it and replace it with a new tank plate. The tank plate cannot be repaired if damaged.
•
Verify that the mixer drive unit is in proper working order prior to attaching to the tank plate.
If the drive unit has been dropped, check for damage. After installation, verify that the mixer
drive is properly mounted to the tank plate. If there is any damage to the drive unit, ensure it is
repaired before using.
•
If, when dropped, the outer magnet was damaged or the outer magnet drive shaft is bent,
attach the drive unit to the tank plate and manually check to determine if there is any contact
between the outer magnet and tank plate. If there is any contact, do not operate the mixer and
have it repaired prior to use.
NovAseptic
®
GMP ATEX Mixer
5