2. Ensure that the supply voltage and
overload protection are set according to
the motor specification data. Ensure that
the electrical cover and the conduit entries
are properly in place. Check that the outer
drive head rotates clockwise.
Drive Unit Thermistor
All standard NovAseptic
®
GMP ATEX drive units are
equipped with two positive temperature coefficient
(PTC) thermistor probes. EMD Millipore recommends
installation of the probes. Thermistor protection
units continuously monitor the temperature of the
motor through a probe embedded in the motor
windings. If the nominal operating temperature of
the probe is reached, a rapid increase in resistance
is converted into a switching function. This can be
used to switch off the motor or to signal a fault.
Accidental breaks in the supply circuits of the
thermistors are also detected.
Motor current higher than recommended could
cause damage to the bearings through decoupling
of the mixing head.
Revolution Counter
!
Observe laws, regulations and/or
standards governing the use, or
intended use, of this equipment.
Any alteration of this equipment
from factory specification may
cause unsafe conditions and will
void the product warranty.
The tachometer must be
connected to an intrinsically safe
barrier.
Do not damage the sensor.
Connection
Connect the socket using the following electric
scheme.
N/N0
1/BN
2/BU
L+
L-
Variable-Frequency Drive
Sine Wave
Power
Operator
Interface
VDF System
Power Conversion
Variable
Frequency
Power
Variable
Frequency
Controller
Mechanical
Power
AC Motor
1540
Power Conversion
A variable-frequency, adjustable-speed drive
(VFD) controls the rotational speed of an
alternating current (AC) electric motor by
controlling the frequency of the electrical power
supplied to the motor. VFDs are also known as
adjustable-frequency drives (AFD), variable
speed drives (VSD), AC drives and inverter
drives.
When starting a motor, a VFD initially applies a
low frequency and voltage to the motor.
A VFD increases the applied frequency and
voltage at a controlled rate and accelerates the
load without drawing excessive current.
14
Installation and User Guide