ID
Area
Reconditioning instructions
2.4
Contacts
Check that cables and connectors are intact and properly
secured. Check that there is no risk of crushing cables and
connectors, and that all cables are fastened with cable ties.
2.5
Batteries
Check the batteries’ voltage and that there is not too great a
difference between the batteries’ charge. A large difference
in charging may indicate that a battery cell is defective. See
installation instructions 'Battery replacement 24V1G'. Keep
in mind that the life expectancy of the batteries is 5 years
when maintained according to these instructions for use.
This means that it may be time to replace them depending
on when the reconditioning is complete and what remedial
actions have been taken on the device in the past.
2.6
Electric seat tilt
Check that the actuator is not visibly damaged. Check the
function of the actuator’s moving parts. Check that all locking
screws are tightened to avoid any play in the end position.
2.7
Electric back recliner
Check that the actuator is not visibly damaged. Check the
function of the actuator’s moving parts. Check that all locking
screws are tightened to avoid any play in the end position.
2.8
Electric brake
Check that the actuator is not visibly damaged. Check the
function of the actuator’s moving parts. Check that all
locking screws are tightened to avoid any play in the end
position. Check that the attachments are intact and securely
fixed to the base.
3 Electric lifting mechanism
3.1
Noise
Listen for noise in the lifting mechanism. Replace the lifting
mechanism in case of noise. Load the device and check that
the lifting mechanism does not slip when starting from the
bottom position. Defective lifting mechanisms can be sent
to Mercado Medic for repair.
3.2
Rotation lock
Make sure that the lifting mechanism does not have any
play when rotating or loading from side to side. Replace
the lifting mechanism in case of play. Defective lifting
mechanisms can be sent to Mercado Medic for repair.
3.3
Lifting columns
Raise the lifting mechanism to the highest position. Wipe
the lifting column of dust and dirt with a dry cloth. Then
lubricate the lifting column using a thin layer of Teflon or
silicone-based grease.
3.4
Attachment
Check that the seat is facing straight ahead relative to the
base. Check that the lifting mechanism attachments to the
seat frame are correctly assembled and tightened. Replace
screws that have damaged key handles or threads.
3.5
Cabling
Inspect cables for any signs of wear, pinching or crushing.
42
Instructions for use for REAL 8000 PLUS Ophthalmology medical clinic chairs
Servicing & reconditioning
21-08997-EN-00