8.5.2 Span Adjustment
Access the span calibration screen as described in Section 6, Local Operation.
Apply a known pressure to the pressure port of the instrument to be calibrated equal to,
or slightly less than the range of the instrument. The applied pressure must be greater
than 50% of the span or span adjustments will not be accepted. Adjust the measured
pressure on the span calibration screen to match the true applied pressure. Press the
♦
(enter) key to store the span offset data.