Getting Started
Page 29
2.5
Connecting and Starting
2.5.1
Safety Instructions for Connection
Adhere to the following safety instructions before you connect the MH70S:
2.5.2
Connecting an Earthing Cable
The MH70S features an earthing connection.
Carry out the following steps to connect the earthing cable:
»
Use an earthing cable with a cross section of at least 0.75 mm
²
.
»
Fasten the cable at the contact point of the earthing connection using a suitable
attachment clamp, as indicated in the following illustration.
Works on the computer system may only be carried out by personnel
qualified for the specific task, who, based on their training and
experience, are able to identify risks and avoid potential hazards when
working with these products/systems.
Any power input must be protected by an external fuse as defined in
chapter
Connecting the Power Supply
.
Implement a readily accessible disconnect device external to the
system.
Read all labels and warnings on the equipment carefully; especially the
power configuration and coding labeling on the system chassis.
Make sure that the voltage supplied by the external power supply
conforms with the voltage given on the power configuration and coding
labeling.
Adhere to the wiring rules valid in your country when connecting the
system.
Connect the earthing cable before making any other connections!
A protective earth connection is essential for the system to meet its EMC
specifications.
When disassembling the system, always detach the earthing cable last.
Earthing
cable
Aachment
clamp
Contact
point