2.2 Installation options (special equipment)
The oven can be set up on the floor or on a table (bench),
model 500 - 700 on a subframe. Ensure at least 150 mm
clearance between back of oven and wall. The distance
between ceiling and oven should not be less than 200 mm
and the minimum wall spacing to the sides is 80 mm.
Generally, a sufficient circulation of air around the oven is
required.
After the oven has been set up horizontally the door can be
adjusted if required (
see chapter -
MAINTENANCE
).
The model 800 is movable. The front castors are lockable by means of a fixation. In order
to assure stability, the front castors
must always be adjusted to the front side
of the
oven for locking.
2.3 Wall mounting (special equipment)
Up to model 700 for all ovens metal mounting brackets are
available. A mounting bracket is delivered with an
incombustible plate. The dimensions of the fixing screws
are to be choosen depending on the total weight (equipment
and loading) of the charge and the quality of the wall.
Pitches of holes for wall fixation (wall bracket)
model A
B
C
mm inch mm inch mm inch
200 489
19,25
770
30,32
300 569 22,4 770
30.32
400 489
19,25
850
33.47
500 649
25,55
930
36,61
600 889
35,00
1090
42,91
540
21,26
700 1129
44,45
1250
49,21 410 16,14
model 600
model 700
Do not place the steriliser on a readily inflammable support!