set
0
I
power
fresh air
0
1
2
3
4
5
6
alarm set
°C
UM 200-800/SM 200-400, ULM/SLM 400-800
3
2 Technical data summary
MEMMERT appliances are electrically heated and thermostatically controlled. Series
UM/SM have natural air circulation whilst series ULM/SLM have fan assisted circulation.
A PT100 DIN B, DIN 43 760 measures the temperature in the working chamber.
Ambient
conditions
Ambient temperature 5
°
C - 40
°
C, rH max. 80%.
Overvoltage Category:
ΙΙ
Contamination degree: 2 according to IEC 664
Setting tempe-
rature range
30
°
C to nominal temperature
(See rating plate)
Working tem-
perature range
From 5°C above ambient to nominal temperature = maximum
temperature (see rating plate)
When the fan of the models ULM / SLM is switched on, the working
temperature can be set from 10°C degrees above ambient up to the
nominal temperature = maximum temperature (according to rating plate).
Overheat
safety device
Serial adjustable overheat control in compliance with DIN 12880 Class 3.1;
Class 2 protection (adjustable temperature limiter) available on request.
Class Aim
of
protection
Scope of
protection
Safety
device
Safety
measures
2
No risk liable to
originate from the
oven in the event
of a fault
TWB
(adjustable temperaure limiter)
automatically shuts off the
oven if the user selected
temperature limit is exceeded.
3.1
Protection of
the oven, the
environment
and the
contents of the
oven
The contents of
the oven are
protected against
overheating
TWW
(adjustable overheat controller)
takes over control of oven if
temperature is exceeded.
Special safety
measures required
depending on the
purpose for which
the oven is used
2.1 Electrical
supply
50 or 60 Hz. Adequate insulated supply which incorporates earth conductor according to
EN 61010, Protection IP 20, no humidity protection according to DIN 40 050. Interference
suppression Grade N according to VDE 0875, limiting values of the class B, part one.
A fuse 250V/15A quick is used for oven protection (in ovens with 400V 3N~ power supply
3 fuses).
Before current-connection please compare label on the oven and instructions of
your local current supplier.
(e.g. in Germany, DIN VDE 0100 with FI - safety device).
Содержание SLM 400