TTechnical description according to EN 60601-1-2
Recommended separation distances between portable and mobile
RF communications equipment and the Memmert steriliser type S..
The steriliser type S.. is intended for use in electromagnetic enviroment in which radiated RF disturbances
are controlled. The customer or of the steriliser type S.. can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the steriliser type S.. as recommended below, according to the maximum power of the
communications equipment.
Rated maximum output
power of trtansmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2
P
80 MHz to 800 MHz
d = 1,2
P
800 Mhz to 2,5 GHz
d = 2,3
P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, whre
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and poeple.
Listing of cables and maximum length of cables
Description of terminal
Type of cable
Maximum length of cable
m
LAN port
RJ45 CAT 6
2
Warning! The use of other cables may result in increased emissions or
decreased immunity of the Memmert steriliser type S..