43
Menu functions
4. Save the setting by pressing the
confirmation key.
Balance
Unit
IP address
-20 %
255.145.136.225
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode
7.3.6
Balance
Description
For appliances of the size 55 and above, application-specific correction of the heat output
distribution (balance) between the upper and lower heating groups is possible. The
adjustment range is from –50 % to +50 %.
upper heat output
upper heat output
lower heat output
-20%
lower heat output
+30%
SETUP
Balance
Unit
IP address
-20
%
192.168.100.100
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode
SETUP
Balance
Unit
IP address
+30
%
192.168.100.100
Subnetmask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer Mode
Fig. 24
Heat output distribution (example): The –20 % (left) setting causes the lower heating
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
default heat output distribution.