24
Operation and control
5.4.2
Basic operation
In general, all settings are made according to the following pattern:
1. Activate the desired parameter (e.g. tem-
perature). To do so, press the correspond-
ing activation key on the left or right or
the respective display. The activated dis-
play is lined in colour, the other displays
are dimmed. The set value is highlighted
in colour.
.5°C
100
TEMP
TIMER
--
--
m
h
22
.4°C
TEMP
37
.0
°C
Set
2. By turning the turn control to the left
or right, adjust the set value (e.g. to
180.0
º
C).
22
.4°C
TEMP
180
.0
°C
Set
3. Save the set value by pressing the confir-
mation key.
The display returns to normal and the
appliance begins adjusting to the defined
set value.
23
.2°C
TEMP
Set 180
.0°C
Additional parameters (air flap position etc.) can be set accordingly.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati-
cally returns to the main menu and restores the former values.
If you want to cancel the setting procedure, press
the activation key on the left or right of the display that
you want to exit. The appliance restores the former
values. Only the settings that you have confirmed by
pressing the confirmation key before cancelling the
setting procedure are accepted.
5.4.3
Operating modes
The appliance can be operated in three modes:
►
Manual mode: The appliance runs in permanent operation at the values set on the
ControlCOCKPIT. Operation in this mode is described in chapter 5.4.4.
►
Timer operation: The appliance runs at the values set until the timer has elapsed. Opera-
tion in this mode is described in chapter 5.4.5 .
►
Programme mode: The appliance automatically runs programme sequences which have
been defined using AtmoCONTROL software at a computer / laptop and then transferred
to the appliance from a USB stick or via Ethernet. Operation in this mode is described in
chapter 5.4.6.