Menu mode

If humidity/room temperature are particularly high, it is possible that the factory setting for 

defrosting, 12 hours, is not sufficient. If this is the case, you should set a more frequent de-

frosting interval, e.g. every 6 hours.
Automatic defrosting is disabled with the setting 

Off

. When operating at low temperatures, 

this over time causes the cooling unit to ice over. Regular defrosting needs to be carried out in 

order to prevent damage to the cooling system. 

7.3.7  

Remote control 

In the setup entry 

Remote control

, you can set whether the 

appliance should be controlled via remote control and if so, 

in which mode. These settings are available:

 

Off

 

Read Only

 

Write

 

Write+Alarm

When the appliance is in remote control mode, the 

 symbol appears in the temperature display. In the 

settings 

Write

 and 

Write+Alarm

, the appliance cannot be 

controlled at the ControlCOCKPIT until the remote control 

has been switched off (setting Off) or set to 

Read Only

.

In order to use the remote control function, program-

ming skills and special libraries are required.

7.3.8  

Gateway

The setup entry 

Gateway

 is used to connect two networks 

with different protocols.
The gateway is set the same way as the IP address (see page 

45).

7.4  

Date and Time

In the 

Time

 display, you can set date and  time, time zone and daylight saving time.

Always set the time zone (and summer time yes/no) before you set the date and time. 

Avoid changing the set time after that since this can lead to gaps or overlapping when 

recording measured values. If you still need to change the time, you should not run a 

programme immediately before or after doing so.

1.  Activate the time setting. To do so, press 

the activation key on the right side of 

the 

Time

 display. The display is enlarged 

and the first adjustment option (

Date

automatically highlighted. 

Date and time

Time
Time zone
Daylight savings

12

.

05

.

2012

12:00
GMT 01:00

Date

Remote Control

   

Off

Gateway                192.168.5  .1

23

.2°C

TEMP

Set 38

.0°C

Remote Control  Off

Gateway  

              

192.168.5  .1

Содержание ICH 110

Страница 1: ...www memmert com www atmosafe net CLIMATE CHAMBER ICH ICH L 100 ATMOSAFE MADE IN GERMANY OPERATING INSTRUCTIONS ICH ICH L...

Страница 2: ...79 E mail service memmert com When contacting customer service always quote the product serial number on the nameplate see page 13 Shipping address for repairs Memmert GmbH Co KG Kundenservice Willi M...

Страница 3: ...vidual configurations and sizes illustrations in this manual may be slightly different from the actual appearance Function and operation are identical Other documents that have to be observed For oper...

Страница 4: ...and interfaces 11 2 6 Designation nameplate 13 2 7 Technical data 14 2 8 Applied directives and standards 15 2 9 Declaration of conformity 15 2 10 Ambient conditions 16 2 11 Scope of delivery 16 2 12...

Страница 5: ...41 6 3 Power failure 42 7 Menu mode 43 7 1 Overview 43 7 2 Basic operation in menu mode using the example of language selection 44 7 3 Setup 45 7 4 Date and Time 48 7 5 Calibration 50 7 6 Program 54...

Страница 6: ...mebody else could be injured if you do not observe the accompanying safety regulation Caution is used for information that is important for avoiding damage 1 1 2 Signs used Warning signs warning of a...

Страница 7: ...do not form any toxic or explosive vapours when heated up see also chapter Intended use on page 8 Warning If the door is open while the appliance is in operation the appliance may overheat and pose a...

Страница 8: ...le for ensuring that personal protective clothing is worn by operating person nel e g work clothes safety shoes and UV safety goggles 1 6 Intended use Climate chambers ICH and ICHL are suitable for st...

Страница 9: ...ge immediately take the appliance out of service and inform your superior You can find information on eliminating malfunctions from page 39 1 9 Switching off the appliance in an emergency Press the ma...

Страница 10: ...ruction 1 ControlCOCKPIT with capacitive function keys and LCD displays see page 26 2 On Off switch see page 23 3 Door handle 4 Chamber fan 5 Castors 6 Cooling unit see page 59 7 Nameplate see page 13...

Страница 11: ...ptimal hygienic properties and corrosion resistance to many but not all chemical compounds caution for example with chlorine compounds The chamber load for the appliance must be carefully checked for...

Страница 12: ...t is located on the right side of the ControlCOCKPIT Fig 4 Ethernet interface Via Ethernet interface the appliance can be connected to a network so that you can transfer pro grammes created with the A...

Страница 13: ...ront of the appliance on the right behind the door see page 10 Typ ICH 260 F Nr Y613 0088 230 V 5 9 A 50 60 Hz 1350 W DIN12880 Kl 3 3 Nenntemp 10 60 C 1 2 3 4 5 6 10 8 9 7 Fig 6 Nameplate example 1 Ty...

Страница 14: ...ing packaging kg 127 209 324 Power W ICH 1350 ICHL 1450 1530 Current consumption A 230 V 50 60 Hz 5 9 230 V 50 60 Hz with light 6 3 6 7 115 V 50 60 Hz 11 8 115 V 50 60 Hz with light 12 6 13 3 max numb...

Страница 15: ...98 EN 61326 A2 2001 EN 61326 A2 2003 Directive 2006 42 EC Directive of the European Parliament and Council dated May 17 2006 on machinery and amending directive 95 16 EC revised Standards complied wit...

Страница 16: ...mentation in the interior and as a consequence could result in short circuits or dam age to electrical parts For this reason sufficient measures to prevent large clouds of dust or aggressive vapours f...

Страница 17: ...sported with a manual pallet jack or forklift truck 110 260 750 Warning You may get your hands or feet squashed when transport ing and installing the appliance Wear protective gloves and safety boots...

Страница 18: ...n edge 3 4 1 Checking for completeness and transport damage Check the delivery note to ensure that the delivery is complete Check the appliance for damage If you notice deviations from the delivery no...

Страница 19: ...pliance see Technical data on page 14 Do not place the appliance on a flamma ble surface Depending on the model see nameplate a 230 V or 115 V power connection must be avail able at the installation s...

Страница 20: ...tion The tilt protection is included in the delivery 1 Screw the tilt protection as shown onto the back of the appliance 2 Bend the tilt protec tion upwards by 90 in the desired distance to the wall c...

Страница 21: ...e to the floor conditions In order to do so every door has two adjuster screws at the top and at the bottom Fig 10 First adjust the door at the top and then if further adjustment is necessary at the b...

Страница 22: ...t it to the power supply Fig 11 Lay the power cable so that it is always accessible and within reach so it can be disconnected quickly in the event of failure or emergencies no one can trip over it it...

Страница 23: ...es fastening material not included in the delivery Fig 12 Water connection Fig 13 Tank holder 4 3 Switching on Switch on the appliance by pressing the On Off switch on the front of the appliance Fig 1...

Страница 24: ...when you touch the interior of the appliance 5 1 Operating personnel The appliance may only be operated by persons who are of legal age and have been instruct ed accordingly Personnel who are to be t...

Страница 25: ...ting capacity the type of drawer used Grid or Shelf must be set in the menu under SETUP see page 55 The chamber must not be loaded too tightly so that proper air circulation in the interior is guarant...

Страница 26: ...rogramme display 9 Activation key for the appliance state 10 Activation key digital backwards counter with target time setting adjustable from 1 minute to 99 days 11 On Off switch 12 Display digital b...

Страница 27: ...tion in this mode is described in chapter 5 4 4 Operation with digital backwards counter with target time setting adjustable from 1 minute to 99 days Timer The appliance runs at the values set until t...

Страница 28: ...00 in steps of 10 FAN 50 FAN FAN Humidity Adjustment range 10 to 80 rh Humidification is indicated by the symbol Dehumidification is indicated by the symbol HUMIDITY 70 rh Set 34 rh A high level of ai...

Страница 29: ...This means operation is only possible at limited temperature and humidity range We recommend a temperature of 25 C and a humidity of 50 RH at most Depending on the ambient conditions there may be som...

Страница 30: ...r activation see page 47 If the timer runs setpoint dependent this is indicated by the symbol in the timer display When the timer has elapsed the display shows 00h 00m All func tions heating etc are s...

Страница 31: ...first description from page 54 3 To start the programme press the confir mation key The programme starts The display shows the programme description in this exam ple Test 012 the programme segment des...

Страница 32: ...atus display then turn the turn control until the hand symbol is highlighted in colour and press the confirmation key 5 5 Monitoring function 5 5 1 Temperature monitoring The appliance is equipped wit...

Страница 33: ...how to adjust temperature monitoring from page 34 please read the description of the individual monitoring functions here Electronic temperature monitoring TWW The manually set monitoring temperature...

Страница 34: ...example The setpoint is changed from 50 C to 25 C section D until the tolerance range of the new temperature setpoint is reached again section E 52 C 48 C ASF active 52 C 48 C ASF alarm ASF active ASF...

Страница 35: ...max C 355 40 0 5 By turning the turn control adjust the desired upper alarm limit value in the example on the right 38 5 C The monitoring temperature must be set sufficiently high above the maximum s...

Страница 36: ...toring 1 Press the activation key to the left of the ALARM display The temperature moni toring setting is automatically activated min 000 C ALARM max 000 C auto off 99K ALARM max 40 0 C auto 0 0K min...

Страница 37: ...urve 5 6 1 Temperature profile 1 Press the activation key to the right of the GRAPH display The display is enlarged and the temperature profile shown 0 4 8 12 16 20 24 C 40 20 60 80 100 Fr 20 10 2010...

Страница 38: ...2010 20 12 09 2012 38 39 40 3 Press the confirmation key to confirm The humidity profile is displayed You can change the display range as described above as well as extend or reduce it 0 4 8 12 16 20...

Страница 39: ...page 55 indicated by the speaker symbol the alarm is additionally signalled by an intermittent acoustic signal If the confirmation key is pressed the acoustic alarm can be tempo rarily switched off un...

Страница 40: ...Error display symbol HUMIDITY Set 55 0 rh rh 55 4 Water tank empty Fill the water tank with dem ineralised distilled water and press the confirmation key page 22 Alarm display MaxAl HUMIDITY MaxAl Set...

Страница 41: ...different Appliance is in wrong mode Change to operating or menu mode by pressing the MENU key Display T E 3 in the temperature display TEMP T E 3 Set 37 0 C 37 4 C Temperature operating sen sor defe...

Страница 42: ...er service Download the firm ware update at mem mert com and install it page 2 page 2 6 3 Power failure In case of a power failure the appliance operates as follows In manual mode After power supply h...

Страница 43: ...IGNALT NE PROG ZEIT UND DATUM CALIB USER ID PROTOCOL LANGUAGE TIME SOUND ON O O ON N N N O ON O 1 2 3 4 5 12 14 15 13 6 7 18 19 8 9 10 11 16 17 Fig 24 ControlCOCKPIT in menu mode 1 Language selection...

Страница 44: ...ngs again press the activation key which you have used to activate the display The appliance returns to the menu overview Only the settings that you have con firmed by pressing the confirmation key be...

Страница 45: ...age 48 If the Setup menu contains more entries than can be displayed this is indicated by the display 1 2 This means that there is a second page of entries To display the hidden entries use the turn c...

Страница 46: ...IP address 255 168 100 100 Subnet mask 255 255 0 0 Alarm temp C F TWW TWB Timer mode 4 Accept the selection by pressing the confirmation key The next three digits of the IP address are automatically...

Страница 47: ...automatic defrosting Due to long term operation with a working temperature below 15 C or due to a damp chamber load and or the door opening frequently ice can form in the interior over time Heavy ici...

Страница 48: ...the remote control has been switched off setting Off or set to Read Only In order to use the remote control function program ming skills and special libraries are required 7 3 8 Gateway The setup ent...

Страница 49: ...savings entry Time 12 05 2012 12 00 GMT 00 00 Date Time zone Daylight savings 6 Accept the selection by pressing the confirmation key The adjustment options are highlighted Time 12 05 2012 12 00 GMT 0...

Страница 50: ...emperature adjustment you will need a calibrated reference measuring device CAL 1 0 5 C CAL 3 1 6 C CAL 2 0 4 C Factory calibration 20 C 30 C 40 C 0 C 10 C Fig 27 Schematic example of temperature adju...

Страница 51: ...adjust the set temperature to 30 C 21 4 C TEMP 30 0 C Set 8 Wait until the appliance reaches the set temperature and displays 30 C The ref erence instrument should display 31 6 C 30 0 C TEMP Set 30 0...

Страница 52: ...the right of the CALIB display The display is enlarged and the temperature adjustment option is automatically highlighted Cal1 C 0 2K Cal2100 0 C 0 1K Cal3180 0 C 0 2K JUSTIEREN Temperatur letzte Jus...

Страница 53: ...y in the appli ance s working chamber 8 Close the door and in manual mode adjust the set humidity to 60 rh HUMIDITY 29 5 rh Set 60 0 rh 9 Wait until the appliance reaches the set humidity and displays...

Страница 54: ...elect function by pressing the confirmation key All programmes available are displayed including the ones saved on the USB storage medium identified by the USB symbol The programme currently available...

Страница 55: ...shown on Sound Keysound At the end If door open On alarm If you want to edit another list entry Turn the turn control until the respective entry e g If door open Special accessories is highlighted in...

Страница 56: ...e is not modified or deleted by reading it out 1 Connect the USB storage medium to the interface on the right side of the control panel 2 Activate the protocol To do so press the activation key on the...

Страница 57: ...AtmoCONTROL manual 7 9 2 User ID activation and deactivation 1 Connect the USB storage medium with the user ID data to the interface on the right side of the control panel 2 Activate the user ID To do...

Страница 58: ...to clean interior prevents build up of material remains that could impair the appearance and functionality of the stainless steel chamber over time The metal surfaces of the appliance can be cleaned w...

Страница 59: ...lower front cover number varies depending on the appliance size and remove the front cover Fig 30 Fig 30 Unscrew and remove the front cover 8 1 5 Peltier dehumidification modules In order to guarante...

Страница 60: ...ing the function of fluores cent tubes Replace the fluorescent tubes only with fluorescent tubes of the same power and type e g UV light tubes only with UV light tubes not daylight tubes For informati...

Страница 61: ...rounding the glass plate Remove the glass plate and put it aside Fig 34 Lay the illumination box on its back and loosen the screws 6 Undo the screw connections on the sockets of the fluorescent tubes...

Страница 62: ...the appliance until they click in place Fig 36 Fig 36 Put the illumination box back in 10 Put in the plug and lock the bayonet catch by turning it right Fig 32 Warning UV light is a danger to your eye...

Страница 63: ...into their national laws It may not be disposed of in normal household waste For disposal please contact your dealer or the manufacturer Any appliances that are infected infectious or contami nated w...

Страница 64: ...2 Explosion protection 8 F Fan speed 28 Fluorescent tubes 60 61 Forklift truck 18 Fresh water container 22 G Graph 37 Grid 47 H Hazards 7 Hot air generator 11 Humidity 28 Humidity adjustment 52 Humidi...

Страница 65: ...adjustment 50 Temperature comparison 50 Temperature deviation 50 Temperature limiter 34 Temperature monitor 32 34 Temperature monitoring 32 Temperature profile 37 Temperature sensor 32 Tilt protection...

Страница 66: ......

Страница 67: ......

Страница 68: ...lli Memmert Stra e 90 96 D 91186 B chenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net Climate chamber ICH D33142...

Отзывы: