32
D33142 | Date 03/2017
Operation and control
3. Press the confirmation key to confirm.
The programme is cancelled.
A cancelled programme cannot be
resumed at the point it was cancelled. It
must be restarted from the beginning.
End
Test 012
10:49
12.Sept.2012
End of programme
The display shows
End
when the programme
is finished.
End
Test 012
10:49
12.Sept.2012
You can now
►
restart the programme as described
►
select another programme for processing in menu mode (see page 54) and run it as
described.
►
return to manual mode. To do so, reactivate it by pressing the
activation key next to the status display, then turn the turn
control until the hand symbol
is highlighted in colour and
press the confirmation key.
5.5
Monitoring function
5.5.1
Temperature monitoring
The appliance is equipped with a multiple overtemperature protection in accordance with
DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in case of a
malfunction:
►
electronic temperature monitoring (TWW)
►
automatic temperature monitor ( ASF)
►
mechanical temperature limiter (TB)
The monitoring temperature of the electronic
temperature monitoring is measured via a separate Pt100 tem-
perature sensor in the interior. Temperature monitoring settings
are made via the
ALARM
display. The settings made apply to all
operating modes.
Manual Mode
12.Sept.2012 13:44
ALARM
max
38
.5°C
2.0
K
min
36
.5°C
auto
+-