Operation and control
10. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
38
.5°C
2.0
K
min
36
.5°C
auto
+-
5.5.2
Humidity monitoring
If humidity monitoring has been triggered,
this is indicated by the humidity display:
the actual humidity is highlighted in red
and a warning symbol
is shown ( Fig.
22 ). If the acoustic alarm has been
activated in the
Sound
menu (see page
55), which is indicated by the speaker
symbol
, the alarm is additionally
signalled by an intermittent acoustic signal.
Information on what to do in this case are
provided in the chapter Malfunctions,
warning and error messages on page
39.
Adjusting humidity monitoring
1. Press the activation key to the left of the
ALARM
display. The temperature moni-
toring setting is automatically activated.
min
000°C
ALARM
max
000
°C
auto off
99K
-
+
ALARM
max
40
.0°C
auto
0.0
K
min
15
.0°C
+-
2. Turn the turn control until the humidity
monitoring entry is highlighted.
ALARM
min
20
.0%rh
max
60
.0%rh
3. Accept the selection by pressing the con-
firmation key. The lower humidity alarm
limit is automatically highlighted.
ALARM
min
20
.0
%rh
max
60
.0%rh
4. By turning the turn control, adjust the
desired lower alarm limit, in the example
on the right 50 % rh.
ALARM
min
50
.0
%rh
max
60
.
0%rh
4
m
min
000°C
A
ON
LIGHT
ON
ON
ON
ON
LIGHT
LIGHT
IGHT
ON
LIGHT
FEUCHTE
30
.0%rh
Set 30.0%rh
0
°C
°C
°C
22
.4
EMP
°C
Set 37.0
L
A
auto
min
35
0
m
3:30
HUMIDITY
Set 70.0%rh
%rh
75
.4
m
FEUCHTE
30
.0%rh
Set 30
.0%rh
C
°C
Set 37.0
m
0
HUMIDITY
Set 70
.0%rh
%rh
75
.4
Fig. 22
Humidity monitoring triggered