D33142 | Date 03/2017
25
Operation and control
Warning!
In case of appliances of a certain size, you can get accidentally
locked in, which is life-threatening. Do not climb into the appliance!
5.3
Loading the appliance
Warning!
When loading the appliance with an unsuitable load, poisonous or
explosive vapours or gases may be produced. This could cause the
appliance to explode, and people could be severely injured or poi-
soned. The appliance may only be loaded with materials which do
not form any toxic or explosive vapours when heated up and cannot
ignite (see also Intended use on page 8). If there is any doubt as
to the composition of materials, they must not be loaded into the
appliance.
Caution:
Check the chamber load for chemical compatibility with the materials of the appliance
(see page 11).
Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the
load capacity are specified in the technical data overview from page 14 .
To achieve the correct heating capacity, the type of drawer used – Grid or Shelf – must be
set in the menu under
SETUP
(see page 55).
The chamber must not be
loaded too tightly, so that
proper air circulation in the
interior is guaranteed. Do
not place any of the chamber
load on the floor, touching
the side walls or right below
the ceiling of the chamber
(Fig. 16, see also the "correct
loading" sticker on the appli-
ance).
In case of improper loading
(chamber loaded too tightly),
the set temperature may
be exceeded or it may take
longer until it is reached.
The appliance is not suitable for long-term storing at sub-zero temperatures. During
permanent operation, the glass door may ice over.
5.4 Operating the appliance
5.4.1
ControlCOCKPIT
In manual operation, the desired parameters are entered at the ControlCOCKPIT on the front
of the appliance (Fig. 17 ). You can also make basic settings here (menu mode). Additionally,
warning messages are displayed, e.g. if the temperature is exceeded. In programme mode,
the parameters defined, the programme description, the programme segment currently active
and programme duration remaining are displayed (for a more detailed description, see page
32).
Fig. 16
Correct placement of the chamber load