Warning message
Cause / Remedy
After starting the program there was a loss of power. The
error report is received when the electricity supply is restored:
Check the mains power supply installation, if no errors can
be found, inform the service agent.
If there is a loss of power when the chamber is under
pressure, then there will be an additional reminder to sterilise
the sterile filter, since this may have become moist and non-
sterile:
Remove the sterile filter at the rear of the autoclave.
Sterilise the filter using the rapid program.
Then replace the filter.
The time limit for the steam generator to reach the necessary
pressure has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
The minimum sterilization temperature has not been reached:
Reduce the size of the load.
If this occurs repeatedly, inform your specialist dealer.
The maximum sterilization temperature has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
Sterilization pressure falls below the minimum level:
Reduce the size of the load.
If this occurs repeatedly, inform your specialist dealer.
The maximum sterilization pressure has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
The difference between measured and theoretical
temperature is too large:
If this occurs repeatedly, inform your specialist dealer.
8
Taking care of your autoclave
8.1
Preparation of instruments
______________________________________________
MELAG - non-rusting materials
All parts of the Vacuklav
®
24-B/L which come into contact with steam are made of non-rusting material. The
autoclave chamber and its door are made of stainless steel, pipes carrying steam are made of Teflon, and
threaded fittings and magnet valves are made of brass.
Drag-in rust
The use of non-rusting materials excludes the formation of rust as a result of the components of the autoclave.
Where rust forms on the autoclave or the items to be sterilized, investigations have repeatedly shown that this
rust has been brought in from other sources. It should be borne in mind that rust can form even on best quality
stainless steel instruments, for example as a result of improper treatment with chemical cleaning agents or
disinfectant during preparation for sterilization.
Preparations of items for sterilization
The example of drag-in rust shows how important it is to prepare items properly for sterilization, and
particularly the following points:
Handpieces and angles should be cleaned before the sterilization in accordance with the manufacturers’
instructions and maintained (e.g. oiled).
E r r o r 3 2 :
P o w e r f a i l u r e
E r r o r 3 3 :
P r e s s u r e d r o p
E r r o r 3 4 :
S t e r i l i z a t i o n T U
E r r o r 3 5 :
S t e r i l i z a t i o n T O
E r r o r 3 6 :
S t e r i l i z a t i o n P U
E r r o r 3 7 :
S t e r i l i z a t i o n P O
E r r o r 3 8 :
S t e r i l i z a t i o n T D
S t e r i l e f i l t e r
s t e r i l i z e