background image

6 Sterilization

15

Storing sterile instruments

The maximum storage time is dependent on the packaging and the storage conditions. For standard-con-
form packaged sterilized equipment – (if protected from dust) it can amount to up to six months. Comply
with the provisions of DIN 58953, part 8 and the criteria specified below for the storage of sterilized equip-
ment:

u

Comply with the maximum storage duration in accordance with the packaging type.

u

Do not store the sterilized equipment in the decontamination room.

u

Store the sterilized equipment in a dust-protected environment e.g. in a closed instrument cabinet.

u

Store the sterilized equipment in a moisture-protected environment (e.g. alcohol, disinfectant).

u

Store the sterilized equipment in an environment protected against excess temperature variations.

Содержание Sterilizer 75

Страница 1: ...shed MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing compe...

Страница 2: ......

Страница 3: ...air sterilizer on and off again 11 Opening and closing the door 12 Changing the mount 12 6 Sterilization 13 Preparing the sterilization material 13 Loading the dry heat sterilizer 13 Information regar...

Страница 4: ...to the device n The power cable may not be replaced by a cable determined to be insufficient Danger of short circuit n Liquids may not be permitted to reach the interior of the device This could resu...

Страница 5: ...perating Manual near to the device It represents a component of the product Formatting rules Example Explanation see Chapter 2 Reference to another text section within this document Symbols used Symbo...

Страница 6: ...the danger of burns In affixing the CE mark the manufacturer declares that this product fulfils the basic requirements of the EMC and low voltage directive The device may not be disposed as domestic...

Страница 7: ...ack Standard tray jack Standard tray cassette unperforated Intended use This hot air sterilizer is intended for use in the cosmetics sector the cosmetic chiropody and the veterinary sector It was desi...

Страница 8: ...amp heating 3 Thermometer 4 Indicator lamp network 5 On Off switch and time setting 6 Door handle 7 Door opens forwards 8 Device foot 9 11 10 Fig 2 Rear view of the device 9 Fan 10 Connection for the...

Страница 9: ...loading versions with devices with trays and standard tray cassettes are possible in the following siz es Tray WxHxD 19 x 2 x 29 cm Standard tray cassette WxHxD 19 x 4 x 29 cm Mounting 1 Mount 1 belo...

Страница 10: ...clearance of 10 cm to other devices and walls especially flammable parts u Ensure sufficient ventilation Ensure sufficient clearance to the top for free removal of the warm air The device may not be u...

Страница 11: ...wise The control lights network and heating will remain illuminated until the time expires The hot air sterilizer can be set to continuous operation In this case the hot air sterilizer must be switche...

Страница 12: ...cedure has started this could result in cooling and the object not being steri lized for a sufficient time 1 Move the On Off switch to the 0 position to switch off 2 Open the door forwards to open 3 M...

Страница 13: ...tion material together in blocks as this will prevent heat equalization u Do not place the sterilization material on cellulose as this can result in a heat build up u Always pack the sterilization mat...

Страница 14: ...zation times table Selecting the sterilization times Do not exceed the specified loading quantities Use only trays and cartridges made of aluminium When needing to pack with foil use aluminium foil Ta...

Страница 15: ...a specified below for the storage of sterilized equip ment u Comply with the maximum storage duration in accordance with the packaging type u Do not store the sterilized equipment in the decontaminati...

Страница 16: ...ng bio indicators thermocouples or maximum thermometers Comply with the valid regional legal specifications When performing the spore test using biological indicators the spore packets are to be clamp...

Страница 17: ...o remove the impurities with this method Only if the chamber mount or chamber seal face has persistent soiling should you use a mild stainless steel cleaning agent with a pH value between 5 and 8 To c...

Страница 18: ...temperature has been set too low The temperature controller rotary knob was unintentionally turned or mal adjusted during unpacking or incorrect handling e g during cleaning Set a higher temperature T...

Страница 19: ...ase observe the maximum load 2 kg including trays Maximum mains voltage supply variations 10 Maximum voltage range 207 253 V Relative humidity up to 31 C max 80 up to 40 C max 50 Degree of protection...

Страница 20: ...cessories 20 11 Accessories Table 2 Accessories Designation of article art no Mounting 1 59030 Mounting 2 59040 Tray 02075 Standard tray cassette 00287 Tray jack 28890 Jack for standard tray cassette...

Страница 21: ...e sterilization material to the required temperature of 180 C Sterilization material Unsterile sterilizable material which is still to be sterilized Sterilized equipment Also referred to as a batch a...

Страница 22: ...0829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Hei luftsteril...

Отзывы: