4 Commissioning
10
4 Commissioning
Requirements of the installation location
CAUTION
Failure to comply with the set-up conditions can result in injuries, malfunctions and/or
damage to the device.
n
Comply with all the specifications of this chapter for initial commissioning.
u
The device is not suitable for operation in explosive atmospheres.
u
The device is only intended for use in interior spaces.
u
The device is conceived for use outside the patient area. The device should be located a minimum of
1.5 m radius away from the treatment area.
u
Install the device in a dry and dust-protected location.
u
Maintain sufficient clearance to the surrounding surfaces in order to ensure sufficient ventilation.
u
Ensure that the sealing device is located away from direct sunshine and outside the range of other
sources of heat.
u
Set-up the device protected against blows or vibrations.
Wall mounting
If the device is not to be placed on a table, it can be mounted on a wall. If this is the case, we recommend
using the wall-mounted roll bracket optionally available. Proceed as follows:
1.
Remove the perforated metal wall-mounting metal panels from the base of the sealing device.
2.
Drill two Ø 6 mm boreholes in the wall with a clearance of 16.5 mm at the desired mounting height.
16.5 cm
3.
Insert two rawl plugs (Ø 6 mm) with round-head screws (Ø 3.5 x 45 mm) in the boreholes.
4.
Hang the sealing device on the screws.