8.2.6  Too much loss of water __________________________________________________________ 

In this case, the water pump runs several times 
after the start of the program. 

Fault diagnosis: 

After a program start, the autoclave ventilates the 
chamber through the flow nozzle which is screwed 
into the measuring block (see also chapter 8.2.1). 
Then steam constantly flows through this jet into the 
condenser container. The resulting loss of water is 
equalized by the amount of water supplied to the 
steam generator at the beginning of a process. If 
water gets lost beyond, this can happen due to 
enlarged flow nozzle, leak in the chamber seal (see 
also 7.9) or at the solenoid valve, the safety valve or 
due to a loose coupling. 
If the steam generator is calcificated due to bad 
water quality, the pump will also start again because 
the overheat protection of the steam generator re-
leases before heat reaches the water inside. 
The pump will start again at release of the steam 
generator's overheat protection and after the end of 
the real pumping time. Thus, the correct supervision 
times for pressure generation and killing may be 
exceeded. 
Call for a service engineer. 
1. If a loud hissing sound can be heard at the 

sliding door while the autoclave is under 
pressure, or if the indicator instruments are 
fogged, then the cause is one of the following: 
either the gasket lip has been turned over by 
improper closing of the sliding door, or the gasket 
for the pressure chamber is defective. 

2. Otherwise, loosen the connection of the hose 

from the pump to the chamber block, bend hose 
into a measuring beaker without kinking it and 
start the pre-warmed apparatus. After 

16 seconds, the pump must stop with an amount 
of raised water between 130 and 170 ml. 
Alternatively, you can measure the weight of the 
condenser container before and after 
sterilization. The increase of weight must be 130-
170g. 

 

If there are deposits of scale on the top side of 
the disassembled hose, then it is necessary to 
clean both this hose as well as the steam 
generator. See Section 8.2.7. 

Fault correction: 

1.  The gasket of the sliding door must be cleaned or 

replaced (see Section 2.3). 

2. If the amount of water is too small, refer to 

Section 1.1.1. 

 

If the amount is correct, optically check all joints 
for leakage when they are under pressure. Watch 
the water discharge hose of the solenoid valve 
(short and straight Teflon hose from solenoid 
valve to a hose-T-junction, left side of the 
apparatus). Here, no steam should flow. Observe 
discharge hose of the safety valve (bent Teflon 
hose from ground plate to hose-T-junction, right 
side of the apparatus). Here, no steam should 
flow. 

 

Tighten all joints, clean or replace solenoid valve 
or clean safety valve (replace inner parts of it). 

  If there are no such leakages, check the flow 

nozzle and perhaps replace it. The diameter of 
the orifice must not be much larger than 0.5 mm. 

 

If the problem is still not solved (but water supply 
is okay, no water loss) and the pressure 
generation operates slower than usual, check the 
steam generator for calcification. It must be 
cleaned or changed. 

8.2.7  Steam generator is clogged with boiler scale ________________________________________ 

In this case, the feed water pump runs several 
times after the start of the program. 

Fault diagnosis: 

The feed water pump (47) starts several times, 
although amount of the water is sufficient (see 
Section 1.1.1), and although there are no leaks (see 
8.2.6). In addition, the upper side of the hose from 

the feed water pump (47) to the chamber block (37) 
is covered with scale. 
You must notify the MELAG Customer Service 
Department. 

Correction of the fault: 

The hose and the steam generator must be cleaned 
or replaced if required. 

8.3 

Fault No. 5 ______________________________________________________________

 

Cause of the fault: 

After the autoclave has once achieved the 
sterilization or disinfection parameter for pressure 
and temperature (as shown by the signal lamp 

RUNNING“ (12) shining continusly), the system is 

not able to maintain these parameters for the 
required period of 4 minutes for sterilization, or 6 
minutes for disinfection, throughout the monitoring 
period of 5.5 minutes for sterilization, or 7.5 minutes 
for disinfection. 

 

Time when the malfunction occurs: 

After the start, and after 5.5 minutes (for sterilization) 
or after 7.5 minutes (for disinfection) after once 
achieving the sterilization or disinfection parameters 
for pressure and temperature (as shown by the 
signal lamp 

RUNNING“ (12) shining continuously). 

The cause here can be as follows: 

Содержание MELAquick 12

Страница 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Страница 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Страница 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Страница 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Страница 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Страница 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Страница 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Страница 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Страница 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Страница 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Страница 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Страница 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Страница 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Страница 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Страница 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Страница 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Страница 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Страница 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Страница 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Страница 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Страница 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Страница 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Страница 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Страница 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Страница 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Страница 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Страница 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...

Отзывы: