MELAG MELAquick 12 Скачать руководство пользователя страница 22

 

 

Page 19 

Fault with warning light 

 

The following hints should give you the possibility to 
correct small faults or to give an exact description of 
the fault to the service engineer. Because the 
MELAquick

®

12

 

 indicates the cause of fault in a 

special way, it is possible to treat faults according to 
the table below. 

In any case, you must report the following data 
to the MELAG Customer Service Department:  

  The point in time when the malfunction was 

reported (for example, during the preheat phase, 
during pressure generation, or during the 

sterilization period).  

  The LED signal which reported the malfunction. 

 The values for pressure and temperature 

indicated at this time. 

 Any unusual developments (for example, 

repeated starting of the water pump). 

The MELAG Customer Service Department requires 
this information in order to quickly repair the 
autoclave at a cost economical to you. 

 

Fault 

No. 

LED 

Fault 

LED 

Door 

LED 

Water 

LED 

Ready 

LED 

Run 

Cause and time of fault 

blinks 

 

 

off 

 

Supervision time for pre-heat exceeded, happens 
during pre-heat phase ("Ready" blinks) 

blinks 

 

 

off 

blinks  Supervision time for pressure generation exceeded, 

happens during pressure generation ("Running" 
blinks) 

blinks 

 

 

off 

blinks  Supervision time for sterilization time exceeded, 

happens during sterilization ("Running" lights) 

on 

blinks 

 

 

 

Attempt to unlock sliding door during process 

11 

on 

 

 

blinks  blinks Maximum steam temperature exceeded after 

autoclave was once "Ready" 

12 

on 

blinks 

 

off 

blinks  The quality (conductivity) of the feed water in the 

storage tank is too bad, appears 2 sec. after the 
“START” button was pressed to start a program. 

15 

Shown only on the print-out 

Power loss during the last cycle 

 
In every of the above cases and in case of "normal" 
misoperation (pressing "start" button when autoclave 
not "READY" or process already running), you will 
hear a beep for 2 seconds. 

On principle, you 

should observe type and time of the fault (e.g. 

during pre-heat phase or during pressure 
generation or during sterilization) and inform the 
service engineer

 in order to make a quick and low-

priced repair possible. 

The autoclave has to be repaired only by the 
manufacturer or by an authorized engineer.

 

8.1 

Fault No. 1 ______________________________________________________________  

Cause of the fault: 

The autoclave's chamber has not reached the 
necessary pre-heat temperature within the 
supervision time of 25 

minutes (sterilization), 

15 minutes (disinfection) respectively. 

Time of fault: 

During pre-heat phase (lamp "READY" blinks) 
Call for a service engineer. 

Diagnosis of fault: 

1.  If the chamber is warm during fault indication and 

the apparatus indicates "READY" after you have 
switched off 

main“ switch and on again shortly 

and waited for a few minutes, there is at least 
one heating cartridge of the pre-heating 
defective. 

2.  If the chamber does not get warm, a fuse has 

failed or a plug contact is defective. 

3.  If the chamber heats up to approximately 150°C 

(temperature of the overheat protection), there is 
a short circuit in the temperature sensor of pre-
heating. 

Fault correction: 

1.  Measure and replace heating elements. 
2. Measure pre-heat output of the electronic and 

replace defective elements. In case of defective 
fuse, you must also check the function of 
electronic exits. The semiconductor-relay should 
only switch if output was directed (e.g. by test 
knob, lower left side of the electronic plate). This 
can be noticed by the respective lamp of the 
output. Otherwise, the relay is defective and the 
electronic print has to be replaced. 

 

If somebody had been working on the autoclave 
before, you should check whether the cable 
shoes of the temperature sensors are sitting on 
the right connector (pre-heat sensor to upper 
connectors). Measure resistance values. 

3.  If there is a short circuit in the temperature entry 

of the pre-heating, the chamber will heat up to 
150°C until overheat protection is released, 
without a light of lamp "READY" because the 
resistance for pre-heat temperature cannot be 
reached. Replace sensor. 

Содержание MELAquick 12

Страница 1: ...facturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testify to the ex ceptional quality of our products which are manufactured exclusively in Germany As all oth...

Страница 2: ...parametric regulation 4 1 5 Printout of the log 5 1 6 Water one way system 5 1 7 No start of program if water quality is inadequate 5 2 Introduction 6 2 1 Preparation of instruments 6 2 2 Formation of...

Страница 3: ...re 17 7 6 Excessive indication of pressure and or temperature 17 7 7 Hissing during run foggy instruments 17 7 8 Sliding door cannot be unlocked 18 7 9 The sliding door cannot be easily moved 18 7 10...

Страница 4: ...rilization mode 2 Locking handle 10 Signal lamp Fault 17 Signal lamp Disinfection mode 3 Water storage tankr 11 Signal lamp Ready 18 Connector for waste water hose 4 Power switch 12 Signal lamp Cycle...

Страница 5: ...Operation manual MELAquick 12 Page 2 1 1 2 Rear View______________________________________________________________________...

Страница 6: ...ection steam generator 41 Steam filter before solenoid valve 49 Solenoid valve for pressure release 34 Temperature sensor pre heating 42 Safety valve 50 Micro switch for door lock 35 Temperature senso...

Страница 7: ...e leaving the factory the autoclave is set to sterilization The operator should select the desired program as soon as possible after switching on the electrical power switch Power on 4 If the autoclav...

Страница 8: ...not only the autoclave but also the instruments will be damaged and become dirty However in daily practice the instruments are prepared insufficiently This is critical especially for the sterilization...

Страница 9: ...n are covered with residues of oil dirt or grease coming form the production 2 2 Formation of rust drag in ________________________________________________ It has already been said that the materials...

Страница 10: ...to successfully verify the sterilisation effectiveness of the autoclave only in conjunction with the sterilization program Test spores which are already packed for example spore strips in paper packi...

Страница 11: ...ck the water level in the container If the water reaches the MAX mark the container must be emptied immediately We recommend to empty the condenser container each time purified water is added to the r...

Страница 12: ...____________________ When the MELAquick 12 is put into operation for the first time it is absolutely necessary to run two consecutive sterilization cycles with no instruments in the autoclave This is...

Страница 13: ...rilze liquids or textiles Observe the instructions above on the reheating and cool down phase if the program is changed between sterilization and disinfection after the autoclave has been preheated 4...

Страница 14: ...ry heating up phase When the required pressure 2 bar and temperature 134 C for sterilization 0 2 bar and 105 C for disinfection is reached the lamp RUNNING shows a continuous light which announces tha...

Страница 15: ...____________________________________ Overheat excluded Overheat during error free process is generally excluded due to the concept of separate steam generation in combination with electronic parametri...

Страница 16: ...__________________________ Because of its parameters control system and its optimised heating the MELAG autoclave MELAquick 12 can be used for sterilization disinfection without interruption So 8 10 s...

Страница 17: ...mpair the autoclave s functional effectiveness Aggressive water pH 5 or 7 can also damage the apparatus It is only for high values of conductivity that the autoclave triggers a warning or fault alarm...

Страница 18: ..._________________________________ If freezing temperatures are possible during transport any water which may remain in the feed water system must be removed in order to prevent damage to the pump and...

Страница 19: ...____________________________ Preheat phase cool down phase During the following phases it is not possible to start a program During the approx 10 15 min preheating phase only once after first switchin...

Страница 20: ...into the steam channel of the chamber from the bottom has to be checked for resistances on plugs and contacts may have to be cleaned or the sensor to be replaced For questions about the precision of...

Страница 21: ...e chamber and allow it to run through Then close the sliding door and blow through the outgoing waste water hose 7 9 The sliding door cannot be easily moved ____________________________________ Normal...

Страница 22: ...time of the fault e g during pre heat phase or during pressure generation or during sterilization and inform the service engineer in order to make a quick and low priced repair possible The autoclave...

Страница 23: ...In this hose and thus in the water discharge hose you can normally see small amounts of flowing steam condensat The presence of prescribed temperature 134 C or 105 C will additionally be indicated by...

Страница 24: ...water reservoir to the steam generator Fault diagnosis If this water dose is too small the pump will start again at release of the steam generator s overheat protection and after the end of the real p...

Страница 25: ...on 2 3 2 If the amount of water is too small refer to Section 1 1 1 If the amount is correct optically check all joints for leakage when they are under pressure Watch the water discharge hose of the s...

Страница 26: ...the clickpoint must be checked The correct clickpoint is as follows when the autoclave is preheated and when the sliding door is locked the microswitch should click approximately halfway through the...

Страница 27: ...e tank with demineralized or distilled water as described in Section 5 3 Then fill again with adequate water see Section 3 3 8 7 Fault No 15____________________________________________________________...

Отзывы: