background image

8

instrument will not perform a test. The high voltage warning indicators

flash, and the beeper sounds, to warn of the hazard until the external

voltage becomes less than 50 V.

When testing, the voltage displayed is the voltage present at the test

terminals of the instrument. If the test voltage is changed, by pressing

either of the test voltage 

s

and 

t

buttons during a test, the new test

voltage is displayed momentarily.

After testing, the voltage displayed is the voltage present on the terminals.

To see what the voltage was immediately prior to the end on the test,

press either test voltage 

s

or 

t

buttons.

Timer indicator

The timer shows minutes and seconds. At the start of a test the timer will

start from zero and at the end of a test it will stop. The duration of the last

test remains on the display until another test is started.

Digital Display

The digital display shows the resistance or current being measured during

a test. The 

/I

button toggles between the two. After a test the display

shows the last measurement made until the timer or voltage test settings

are changed or the test start/stop button is pressed.

Analogue display

This simulates an analogue meter movement to give the user a better

"feel" for how a measurement is progressing. The analogue display shows

resistance only.

Battery bar graph

This is a battery symbol on the LCD display comprising 4 pairs of

segments. The battery is monitored continuously when the instrument is

turned on. The charge remaining in the battery is shown in the table

below.

Fully charged battery

50% charged battery

Tests cannot be started, and the battery may fail

at any time.

Symbol flashes when there is not enough charge 

for a test. The instrument then turns itself off

When mains power is present the indicator shows the battery is being

charged by animating the segments of the bar graph.

Voltage at terminals

The test voltage will default to that selected in the previous test. If the

instrument has been switched off since the previous test, the test voltage

defaults to 250 V.

If there is an external voltage greater than 50 V present, this will be

displayed regardless of changes made to the test voltage. In this case the

Содержание MIT510/2

Страница 1: ...M MIT510 2 5kV Digital Insulation Tester USER MANUAL...

Страница 2: ...d items should be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby ra...

Страница 3: ...ndicate best practice Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA CONTENTS 3 Safety warnings 2 Cleaning 3 Introduction 4...

Страница 4: ...harge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 2 is a compact microprocessor controlled high voltage d...

Страница 5: ...t first will perform a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set...

Страница 6: ...played briefly When the test has stopped the display continues to show the voltage present on the test leads Pressing either test voltage s or t button will then display the test voltage immediately b...

Страница 7: ...rminals marked and G These terminals are designed to accept only the test leads supplied Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock wi...

Страница 8: ...display shows the last measurement made until the timer or voltage test settings are changed or the test start stop button is pressed Analogue display This simulates an analogue meter movement to give...

Страница 9: ...en the positive and negative terminals These leakages have a significant effect and can occur through the air itself If the guard lead is connected to ground then since the negative terminal is at the...

Страница 10: ...rt circuit charge current 3 mA 5 kV Capacitor charge time 3 seconds per F at 3mA to 5 kV Capacitor discharge time 120 ms per F to discharge from 5000 V to 50 V Capacitance measurement 500 V minimum te...

Страница 11: ...nA at all voltages Interference rejection 1 mA rms per 250 V test voltage up to a maximum of 2 mA Timer range Counts up to 99 minutes from start of test Test regimes Auto IR Lead set Three flexible si...

Страница 12: ...insulated small clips 8101 183 3 x 10 m with medium insulated clips 1000 441 3 x 15 m with medium insulated clips 1000 442 3 x 3 m with large insulated clips 6220 811 3 x 10 m with large insulated cl...

Страница 13: ...or your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA n...

Страница 14: ...2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri w Mumbai 400053 Maharashtra India T 91 22 267...

Отзывы: