background image

Capacitance display

At the end of a test the capacitance of the circuit under test is calculated

and displayed. After a test the display shows the last measurement made

until the V

s

, V

t

, or 

TEST

buttons are pressed.

Time constant:

At the end of a test the time constant of the circuit under test is calculated

and displayed. After a test the display shows the last measurement made

until the V

s

, V

t

, or 

TEST

buttons are pressed.

Measurements above 100 G

Measurements up to 100 G

can be made without any special precautions,

assuming that the test leads are reasonably clean and dry. The guard lead

can be used to remove the effects of surface leakage if necessary. When

measuring resistances above 100 G

, the test leads should not be allowed

to touch each other, or any other object since this will introduce leakage

paths. Sharp points at the test lead connections should also be avoided

since this will encourage corona discharge.

The output is isolated, and so will float relative to ground such that the

positive terminal is at plus half of the test voltage, and the negative terminal

is at minus half of the test voltage with respect to ground. Leakages

therefore occur between the positive terminal and ground, between the

negative terminal and ground, and directly between the positive and

negative terminals. These leakages have a significant effect and can occur

through the air itself.

If the guard lead is connected to ground, then since the negative terminal

is at the same voltage as the guard terminal, the leakage into the negative

terminal will be considerably reduced. This will improve accuracy because

the current flowing into the negative terminal is measured by the

instrument and used to calculate resistance. This technique is only

permissible if the item under test is isolated from ground. “Isolated” in

this context means insulated by a resistance of at least 5 M

for the

positive terminal, or at least 10 k

for the negative terminal.

Conversely, if the positive terminal is grounded, then the negative terminal

will be at a voltage equal to the test voltage relative to ground, which will

result in an increase in leakage current, and worsening of measurement

accuracy.

When making measurements above 100 G

therefore, the user should

ground the Guard Lead where possible, otherwise parallel leakage paths

may occur.

Alternatively, screened leads are available as an optional accessory from

Megger. The lead to the negative terminal is fully screened. The screen is

plugged into the Guard terminal, diverting any stray leakage currents. This

considerably improves measurements made with a floating output, where

the leads might touch each other or anything other than the test piece.

9

+ Test V

2

- Test V

2

Ground

+

-

+ Test V

0V

Ground

Guard

+

-

Содержание MIT510/2

Страница 1: ...M MIT510 2 5kV Digital Insulation Tester USER MANUAL...

Страница 2: ...d items should be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby ra...

Страница 3: ...ndicate best practice Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA CONTENTS 3 Safety warnings 2 Cleaning 3 Introduction 4...

Страница 4: ...harge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 2 is a compact microprocessor controlled high voltage d...

Страница 5: ...t first will perform a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set...

Страница 6: ...played briefly When the test has stopped the display continues to show the voltage present on the test leads Pressing either test voltage s or t button will then display the test voltage immediately b...

Страница 7: ...rminals marked and G These terminals are designed to accept only the test leads supplied Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock wi...

Страница 8: ...display shows the last measurement made until the timer or voltage test settings are changed or the test start stop button is pressed Analogue display This simulates an analogue meter movement to give...

Страница 9: ...en the positive and negative terminals These leakages have a significant effect and can occur through the air itself If the guard lead is connected to ground then since the negative terminal is at the...

Страница 10: ...rt circuit charge current 3 mA 5 kV Capacitor charge time 3 seconds per F at 3mA to 5 kV Capacitor discharge time 120 ms per F to discharge from 5000 V to 50 V Capacitance measurement 500 V minimum te...

Страница 11: ...nA at all voltages Interference rejection 1 mA rms per 250 V test voltage up to a maximum of 2 mA Timer range Counts up to 99 minutes from start of test Test regimes Auto IR Lead set Three flexible si...

Страница 12: ...insulated small clips 8101 183 3 x 10 m with medium insulated clips 1000 441 3 x 15 m with medium insulated clips 1000 442 3 x 3 m with large insulated clips 6220 811 3 x 10 m with large insulated cl...

Страница 13: ...or your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA n...

Страница 14: ...2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri w Mumbai 400053 Maharashtra India T 91 22 267...

Отзывы: