background image

Continuity Measurement [

]

Range switch set to [

] range

Test Lead options 1 or 2 

The MFT will automatically display continuity resistance when
connected.

Lead Null 

Short test leads together and press the YELLOW test button on the
instrument or Switched Probe.

Continuity Buzzer (

Z

)

The buzzer will sound automatically when a circuit is made (default
threshold <2 

)

Buzzer Threshold (

Z

)

Select BUZZER range and press the ORANGE button until required
range is displayed.
Options, 2 

, 5 

, 10 

, 20 

, 50 

, and 100 

Warning:

Ensure the circuit to be tested is disconnected and

isolated before starting testing.

Insulation testing  [M

]

Range switch set to [250 V, 500 V or 1 kV] range

Test Lead Options 1 or 2

For insulation voltage range select 250 V, 500 V or 1 kV as required.

Press the YELLOW test button on the instrument or on the
switched probe to activate the test. 

Insulation Test lock

Press the YELLOW test and RED lock buttons together to lock on
the insulation test mode.

The warning triangle is displayed during test lock.

To disable test lock press the YELLOW test button again.

Warning:  

Hazardous voltages may be continuously present on the test
probes when the test button is locked down.

Содержание MFT1500/2

Страница 1: ...features and should not be regarded as a substitute for normal safe working practices Do not move the rotary switch positions while a test is in progress The LCD neon voltage indicators cannot reveal...

Страница 2: ...lectricians should be used Symbols used on the instrument are F Caution risk of electric shock G Caution refer to accompanying notes t Equipment protected throughout by Double Insulation Class II c Eq...

Страница 3: ...100 Warning Ensure the circuit to be tested is disconnected and isolated before starting testing Insulation testing M Range switch set to 250 V 500 V or 1 kV range Test Lead Options 1 or 2 For insulat...

Страница 4: ...ed value may differ to the anticipated value This is caused by noise on the supply Repeating the test may result in a different reading if this is the case refer to page 13 of the handbook for possibl...

Страница 5: ...the RED lead to 1st PHASE Bonded Metalwork Testing Hi Test Lead set OPTION 1 or 2 1 Connect the GREEN lead to the bonded metalwork 2 Connect the RED lead to PHASE 3 Supply voltage is displayed 4 The...

Страница 6: ...Connect the test leads to the RCD as shown above 3 Press the YELLOW test button on the instrument or the switch probe On 1 2I the message 1999 ms is displayed after 2 seconds NOTE The message TrP will...

Страница 7: ...enabled by a long press on the blue degree button As selective RCD test mode is enabled the MFT will emit a long beep and the display will show the exclamation mark symbol To disable Selective RCD mod...

Страница 8: ...it 1 T 1 416 298 6770 Scarborough Ontario M1S 3R2 F 1 416 298 0848 Canada E casales megger com This instrument is manufactured in the United Kingdom The company reserves the right to change the specif...

Отзывы: