Display:
3 digit L.C.D.
Temperature Range:
Operating:
-5˚C to 40˚C (0 - 90% RH non Condensing)
Storage:
-25˚C to 65˚C (0 - 95% RH non Condensing at
4 0 ˚ C )
Thermal Protection:
Thermal cut-out will prevent overheating caused
by repetitive testing on highcurrent range.
Safety: Meets the requirements for double insulation to IEC1010-1(1995)
EN61010 (1995) at 230 V installation Category
III*
, without the need for
separately fused test leads. If required, fused test leads are available as an
optional accessory
. *
Relates to transient overvoltage likely to be found in
fixed installation wiring.
Fuses:
- Internal 10A (F) 440 V 10kA ceramic HBC
- Mains power cord fused plug (when applicable):
10 Amp fuse to BS1362
E.M.C:
In accordance with IEC 61326 including amendment No.1
Note:
Overvoltage spikes may cause a reset to the original
power ‘On’ state before reverting to normal operation.
Environmental Protection:
IP54
Dimensions:
220 mm x 92 mm x 55 mm
Weight:
1200g (including leads, case & battery cells)
Cleaning:
Wipe disconnected instrument with a clean cloth
dampened with soapy water or Isopropyl Alcohol (IPA).
ACCESSORIES
Supplied
Part number
User Guide
6172-087
Power cord test lead with 3 pin plug to BS1363/A
6231-601
or
Power cord test lead with CEE 7/7 plug
6231-593
Dual test lead with probe
6231-591
Test-&-carry case
6420-092
Optional
Test Lead Set red/balck
6220-437
Crocodile clip, black, for use with dual test lead
6280-284
Earth bond test lead with probe and 3 pin plug
6231-586
to BS1363/A
Fused probe and clip set (2 probes and 3 clips)
1000 V max.10A fuse
6180-405
Power cord
The power cord supplied with your
LT7
forms part of the measuring circuit of
the instrument. The overall length of this lead must not be altered.
If the power cord plug is not suitable for your type of socket outlets, do not use
an adaptor. You may change the plug once only by cutting the cord as close
as possible and fitting a suitable plug.
The colour code of the cord is:
Earth (Ground)
Yellow / Green
N e u t r a l
B l u e
Phase (Line)
B r o w n
If using a fused plug, a 10 Amp fuse to BS 1362 should be fitted.
Note: Aplug severed from the power cord should be destroyed, as a plug
with bare conductors is hazardous in a live socket outlet.