36
IDAX 300/322/350
ZP-AG03E
AG033586EE
6 APPENDIX AERROR MESSAGES
6
A
ppendix A
Error messages
6.1 Error messages
No. Message
Explanation
347
Output voltage is not
within specified levels.
The applied voltage differs
from the desired.
361
Overvoltage
An overvoltage occurred.
364
Measured capacitances
don’t match
Disagreement between val-
ues of capacitance measured
for different configurations
365
Specimen capacitance
below limit.
Measured capacitance below
specified limit.
366
Specimen capacitance
above limit.
Measured capacitance above
specified limit.
367
Measured DC current>-
MaxDCCurrent
Measured DC current ex-
ceeds limits set by MaxDC-
Current variable.
368
Measured hum current>-
MaxHumCurrent
Measured interference (hum)
current exceeds limits set by
MaxHumCurrent variable.
(347) Output voltage is not within
specified limits
Measured output voltage is outside specified limits.
Possible reasons and countermeasures
1.
Voltage electrode, Generator, is grounded:
▪
check measurement set-up and disconnect ground
▪
change measurement configuration if terminal of
test object cannot be disconnected from ground
2.
Voltage electrode, Generator, is connected to
measuring electrode, Input or Ground:
▪
check measurement set-up and disconnect
measuring or guard electrodes from the voltage
electrode. Voltage electrode, Generator, must not be
connected to either measuring or guard electrode
3.
High stray capacitances to ground or high capaci-
tance of the test object:
▪
Lower the highest frequency used in measurement
plan (see also Measurement variables)
▪
Lower the InitFrequency in the measurement plan
(see also Measurement variables)
▪
Lower the test voltage
4.
If you try to use an old version of IDAX software,
version 3.2 or earlier but the firmware in IDAX is
for IDAX software 4.0 or newer, the IDAX software
do not understand the incapability and it usually
result in error 347.
▪
Please check IDAX software and if version 3.2 or
earlier upgrade to 4.0 or newer (this new software
will automatically upgrade firmware if necessary).
(361) Overvoltage
Measurement aborted due to overvoltage detected on
measuring electrode, Lo. Potential difference between
signal ground and real ground exceeds limiting value.
Possible reasons and countermeasures
1.
Ground electrode is not connected to true ground:
▪
connect Ground electrode to true station/substation
ground
2.
Transients caused by accidentally disconnected
ground connection:
▪
check ground connection
(364) Measured capacitances don’t
match
Values of capacitance measured for different
configurations, the uST, GST-Guard and GSTGround
are in disagreement.
Possible reasons and countermeasures
1.
When performing UST measurement, measuring
electrode, Lo, is connected together with Ground
electrode or Lo is connected to ground:
▪
check measurement set-up and make sure that
measuring electrode, Lo, is connected to a
nongrounded terminal of the test object and that
Ground electrode is connected to ground.
(365) Specimen capacitance below
limit
Measured capacitance below limit specified in C-file
by MinSpecimenC.
Possible reasons and countermeasures
1.
Measured capacitance higher than 10 pF.
Specimen size is very small which results in low
value of capacitance:
▪
change limit set by MinSpecimenC to an
approximately 10% lower value than measured
capacitance
Содержание IDAX 300
Страница 2: ......
Страница 5: ...AG033586EE ZP AG03E IDAX 300 322 350 5...
Страница 11: ...AG033586EE ZP AG03E IDAX 300 322 350 11 2 Safety...
Страница 19: ...AG033586EE ZP AG03E IDAX 300 322 350 19 3 Instrument description...
Страница 33: ...AG033586EE ZP AG03E IDAX 300 322 350 33 4 IDAX 5 Software...
Страница 35: ...AG033586EE ZP AG03E IDAX 300 322 350 35 5 Calibration...
Страница 43: ...AG033586EE ZP AG03E IDAX 300 322 350 43...
Страница 45: ...AG033586EE ZP AG03E IDAX 300 322 350 45 8 Specifications...