OPERATION

9

storing results. The main menu screen will also display “

300 MEMORY 

FULL

” instead of the number of stored tests.

Data is stored for up to 10 years in battery backed 

RAM

.

CURRENT Menu

There are two options, 

SET

, and 

LIMIT

.

Set

This option allows you to set the desired test current. Simply enter the 
desired value using the numeric keys on the keyboard and press Enter.  If 
the current selected is greater than the limit that has been set, DLRO200 
will beep, cancel your entry in the Set Current field and wait for you to 
enter a valid current.

Limit

Some samples to be tested may not be able to withstand the passing of 
heavy currents. In this case set a maximum level of test current to prevent 
the accidental entry of an excessive test current. This level defaults to 
200A. If a lower limit is required enter it using the keyboard and press 
Enter. If the 

LIMIT

 current selected is lower than the 

SET

 current, the 

SET

 current will be reduced to the same value as the 

LIMIT

.

The Main Menu screen will show the active Current Limit beneath the Set 
Current.

Warning Messages

The display will, from time to time, display certain warning messages, 
which may affect measurement accuracy or operator safety.

Test Circuit Continuity

A good measurement requires both the current carrying circuit and the 
voltage detection circuit to be completed by the item under test. The 
DLRO200 checks for this continuity after the 

TEST

 button has been 

pressed.

Initially the Potential circuit is checked. If there is poor continuity in this 
circuit a message will appear on the display stating “

POTENTIAL FAIL

”.  

Press Enter to clear this message and correct the discontinuity in the P 
circuit.

Press 

TEST

 again. If the P circuit is complete DLRO200 will attempt to 

pass the test current.  If the C circuit continuity is inadequate, after a little 
while, DLRO200 will display the message “

CURRENT LOOP FAIL

”.  

Press Enter to clear the message. Correct the fault and start the test again.

External Voltage Warning

The item being tested must be voltage free. If, at any time while DLRO200 
is connected, the item being tested has a voltage in excess of 10 volts ac 
peak or dc with respect to the instrument ground potential, a message 

EXTERNAL VOLTAGE ON TERMINALS

” will appear on the display. 

This is a warning that the item under test is live and might be dangerous. 
A test cannot be performed in this condition.

Remove the external voltage. The display will revert to the main menu 
screen.

If the voltage was detected at the start of a test you may now start the test. 
If the voltage was detected during or at the end of a test, the results will 
be invalid, will not be stored and the test will need to be repeated after 
removing the source of voltage.  

Discharge Current Warning

A message 

CURRENT FLOW

 will appear if a current greater than 

approximately 10 mA is still flowing after a test is completed. This suggests 
that an inductive load has been accidentally tested and is still discharging. 
Do not disconnect the current loop while the discharge warning is showing.

www.actoolsupply.com

www.actoolsupply.com

Содержание DLRO 200

Страница 1: ...M DLRO 200 High Current Low Resistance Ohmmeter USER MANUAL www actoolsupply com www actoolsupply com...

Страница 2: ...a voltage upon it greater than 10 volts ac peak or dc with reference to ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out n Testing inductive circui...

Страница 3: ...tives CONTENTS 3 SAFETY WARNINGS 2 GENERAL DESCRIPTION 4 OPERATION 5 Main Menu Screen 5 TEST Menu 6 OPTIONS Menu 7 CURRENT Menu 9 Warning Messages 9 Entering Notes In The Memo Screen 9 Test Result Sto...

Страница 4: ...nt All four terminals are floating A warning will be given if dangerous voltages are connected to any terminal C or P with respect to ground GENERAL DESCRIPTION 4 The Megger DLRO 200 is an a c powered...

Страница 5: ...LRO200 will beep and clear the SET CURRENT field Re enter the desired current if necessary increasing the current limit to a value equal to or greater than the desired level See separate topic for set...

Страница 6: ...ely connected to the item under test before pressing the Test button The DLRO200 checks voltage and current loop continuity as in Normal Mode and if adequate passes a continuous d c current repeating...

Страница 7: ...sing the two axis cursor control and the Enter key These are Retrieve Passbands Set Clock Delete Data and Storage Retrieve allows the recall of results which have been stored within the DLRO200 s inte...

Страница 8: ...ng may continue without Passbands The passband option allows you to set upper and lower limits between which the test result must lie if it is to be assigned a Pass Readings that fall above the upper...

Страница 9: ...sed Initially the Potential circuit is checked If there is poor continuity in this circuit a message will appear on the display stating POTENTIAL FAIL Press Enter to clear this message and correct the...

Страница 10: ...u do not wish to add notes press the Test button and a new test will be started or the Enter button to return to the Main Screen TEST RESULT STORAGE Each test is numbered sequentially dated and timed...

Страница 11: ...r these leads should be adequately insulated and sufficiently strong mechanically to support their own weight if long leads are being used The size of the current leads will limit the maximum resistan...

Страница 12: ...ent Ranges Full Scale Resolution Volts full scale Max Test Current 999 9 0 1 0 2 V 200 A 9 999 m 1 2 V 200 A 99 99 m 10 4 V 40 A 999 9 m 100 4 V 4 A Subject to supply voltage conditions see below Accu...

Страница 13: ...M Megger is a registered trademark DLRO200_UG_en_V04 0810 www actoolsupply com www actoolsupply com...

Отзывы: