Megger DLRO 200 Скачать руководство пользователя страница 8

OPERATION

8

displayed.

Use the cursor control Up and Down arrows to adjust the highlighted data. 
Step to the next item by using the Right cursor control arrow.

The row below 

DD MM YY HH MM

 contains respectively the date, the 

month, the two-digit year (21st century is assumed), the hour of the day in 
24-hour notation and the minute. These must be entered in this sequence 
regardless of the date format you wish to use.

The lower 

DD/MM/YY 

shows the current date format. Pressing the 

cursor control Up arrow will cycle through the available options 

DD/

MM/YY, MM/DD/YY

 or 

YY/MM/DD

. This is updated when you press 

Enter to exit the Set Clock function. However, tests already stored prior to 
changing the date format will retain the old format.

Delete Data

Select Delete Data if you wish to clear DLRO200’s memory of stored data. 
In case you have selected this Option by accident you will be asked to 
confirm that you wish to delete the data. The default is 

NO

.  Change this 

to 

YES

 and press Enter if you wish to delete all data. 

Please

 

note - ALL 

stored data will be deleted.

Storage

The Storage option sets the instrument to Store or No Store. At the end of 
each test you will be offered the opportunity to change this setting for the 
test just completed and subsequent tests.

At the end of each test you may enter notes into the 

MEMO

 screen, which 

is accessed by pressing any key on the alphanumeric keyboard.  This will 
also cause the test to be automatically stored regardless of other settings.

If all memory has been filled, a message 

MEMORY FULL

 will appear and 

no more tests results will be stored although testing may continue without 

Passbands

The passband option allows you to set upper and lower limits between 
which the test result must lie if it is to be assigned a Pass. Readings that fall 
above the upper limit or below the lower limit will be assigned a Fail.

The upper and lower limits are entered via the keyboard complete with 
decimal points, where applicable, and including the m or µ symbol as 
appropriate. It is not necessary to enter the 

Ω

 symbol. Entering the m or µ 

symbol tells DLRO200 that the entry in that particular field is complete and 
the cursor will therefore move to the next field. The upper limit must be 
less than or equal to 999.9 m

Ω

 and the lower limit must be less than the 

upper limit.

After completing the upper and lower limits the cursor will move to the 

ENABLE

 or 

DISABLE

 options. Highlight your choice using the right and 

left controls and press Enter. You will return to the Main Menu screen.

Note: Passbands will remain enabled or disabled until you re-enter 
this screen and change the selection. If all that is required is to change 
Passbands from 

ENABLED

 to 

DISABLE

 or vice versa, enter the 

PASSBAND

 screen and press Enter until the 

ENABLED / DISABLED

 

choice is shown at which time you may change the selection. Pressing 
Enter will skip over the limit values without changing them if none of the 
numeric keys has been pressed.

If passbands have been set and enabled, at the end of a test the display will 
show the results and will show the word 

PASS

 or 

FAIL

 as appropriate. A 

PASS

 will also be signified by a short beep from the instrument, while a 

FAIL

 will be signified by a long beep.

Set Clock

This option sets the date and time as well as setting the date format.  
When you enter this screen the current date, time and date format will be 

www.actoolsupply.com

www.actoolsupply.com

Содержание DLRO 200

Страница 1: ...M DLRO 200 High Current Low Resistance Ohmmeter USER MANUAL www actoolsupply com www actoolsupply com...

Страница 2: ...a voltage upon it greater than 10 volts ac peak or dc with reference to ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out n Testing inductive circui...

Страница 3: ...tives CONTENTS 3 SAFETY WARNINGS 2 GENERAL DESCRIPTION 4 OPERATION 5 Main Menu Screen 5 TEST Menu 6 OPTIONS Menu 7 CURRENT Menu 9 Warning Messages 9 Entering Notes In The Memo Screen 9 Test Result Sto...

Страница 4: ...nt All four terminals are floating A warning will be given if dangerous voltages are connected to any terminal C or P with respect to ground GENERAL DESCRIPTION 4 The Megger DLRO 200 is an a c powered...

Страница 5: ...LRO200 will beep and clear the SET CURRENT field Re enter the desired current if necessary increasing the current limit to a value equal to or greater than the desired level See separate topic for set...

Страница 6: ...ely connected to the item under test before pressing the Test button The DLRO200 checks voltage and current loop continuity as in Normal Mode and if adequate passes a continuous d c current repeating...

Страница 7: ...sing the two axis cursor control and the Enter key These are Retrieve Passbands Set Clock Delete Data and Storage Retrieve allows the recall of results which have been stored within the DLRO200 s inte...

Страница 8: ...ng may continue without Passbands The passband option allows you to set upper and lower limits between which the test result must lie if it is to be assigned a Pass Readings that fall above the upper...

Страница 9: ...sed Initially the Potential circuit is checked If there is poor continuity in this circuit a message will appear on the display stating POTENTIAL FAIL Press Enter to clear this message and correct the...

Страница 10: ...u do not wish to add notes press the Test button and a new test will be started or the Enter button to return to the Main Screen TEST RESULT STORAGE Each test is numbered sequentially dated and timed...

Страница 11: ...r these leads should be adequately insulated and sufficiently strong mechanically to support their own weight if long leads are being used The size of the current leads will limit the maximum resistan...

Страница 12: ...ent Ranges Full Scale Resolution Volts full scale Max Test Current 999 9 0 1 0 2 V 200 A 9 999 m 1 2 V 200 A 99 99 m 10 4 V 40 A 999 9 m 100 4 V 4 A Subject to supply voltage conditions see below Accu...

Страница 13: ...M Megger is a registered trademark DLRO200_UG_en_V04 0810 www actoolsupply com www actoolsupply com...

Отзывы: