Megger DLRO 200 Скачать руководство пользователя страница 12

Warnings: 

 

Current flowing LED, Inadequate ‘C’  

 

 

and ‘P’ contact, External Voltage,    

 

 

Memory Full, Overheating.

Data: 

Transfer: 

Real time or batch download via  

 

 

 

RS232 using Download Manager  

 

 

 

(supplied free of charge).

 

Storage  

300 sets of results and memo  

 

 

Capacity: 

battery backed 10 years.

 

Memo field: 160 characters max.

Test Current: 

Range: 

10 A to 200 A* unsmoothed d.c. 

 

Accuracy:  ±2% ±2 A

Voltmeter input impedance: 

> 200 k

Ω

Hum rejection: 

 

10 V rms

Power Supply: 

DLRO200:  100 - 265 V a.c. rms 50/60Hz  

                                DLRO200-115:  100 - 130 V a.c. rms 50/60Hz
Input Current: 

 

10 A maximum

Temperature: 

Operation:  -10

o

C to +50

o

C

 

Storage: 

-25

o

C to +65

o

C

 

Calibration:  20

o

C

 

Co-efficient: < 0.05% per 

o

C

Humidity (max): 

 

95% RH non-condensing

Altitude (max): 

 

2000 m

Safety: 

 

IEC61010-1

EMC: 

 

EN61326-1

Operational inaccuracies: 

Refer to www.megger.com

Dimensions: 

 

410 x 250 x 250 mm

Weight: 

 

< 15 kg.

 

Measurement Ranges
Full Scale  Resolution 

Volts (full scale) 

Max Test Current*

999.9 µ

Ω

  0.1 µ

Ω

 

 

0.2 V 

 

200 A

9.999 m

Ω

 1 µ

Ω

 

 

2 V 

 

200 A

99.99 m

Ω

 10 µ

Ω

 

 

4 V 

 

40 A

999.9 m

Ω

 100 µ

Ω

 

 

4 V 

 

4 A

*  Subject to supply voltage conditions (see below).
Accuracy is better than ±1% from 300 µ

Ω

 to 100 m

Ω

 and ±2% from 

100 µ

Ω

  to 300 µ

Ω

, if the maximum possible current is used. If lower 

currents are used, the following table applies:
Current  Error
2 A* 

5.6% + 100 µ

Ω

5 A* 

2.6% + 40 µ

Ω

10 A 

1.6% + 20 µ

Ω

20 A 

1.2% + 10 µ

Ω

50 A 

0.8% + 4 µ

Ω

100 A 

0.7% + 2 µ

Ω

200 A 

0.7% + 1 µ

Ω

*It is not possible to select a current less than 10 A but since the open 
circuit terminal voltage is somewhere between 2 V and 5 V (if input power 
is 100 V to 250 V), actual test current can be as low as 2 A when loop 
resistance is 1 

Ω

. If actual current is less than 2 A a ‘Current loop fail’ will 

be reported. 
Measurement: 

Modes: 

 

Manual, Auto, Continuous.

 

Test Time:   

< 10 seconds NORMAL or  

 

 

 

 

AUTO mode, Refreshed every    

 

 

 

2.4s in CONTINUOUS mode

Display: 

Measurement:  Large, high resolution backlit    

 

 

 

liquid crystal display

 

SPECIFICATIONS

12

www.actoolsupply.com

www.actoolsupply.com

Содержание DLRO 200

Страница 1: ...M DLRO 200 High Current Low Resistance Ohmmeter USER MANUAL www actoolsupply com www actoolsupply com...

Страница 2: ...a voltage upon it greater than 10 volts ac peak or dc with reference to ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out n Testing inductive circui...

Страница 3: ...tives CONTENTS 3 SAFETY WARNINGS 2 GENERAL DESCRIPTION 4 OPERATION 5 Main Menu Screen 5 TEST Menu 6 OPTIONS Menu 7 CURRENT Menu 9 Warning Messages 9 Entering Notes In The Memo Screen 9 Test Result Sto...

Страница 4: ...nt All four terminals are floating A warning will be given if dangerous voltages are connected to any terminal C or P with respect to ground GENERAL DESCRIPTION 4 The Megger DLRO 200 is an a c powered...

Страница 5: ...LRO200 will beep and clear the SET CURRENT field Re enter the desired current if necessary increasing the current limit to a value equal to or greater than the desired level See separate topic for set...

Страница 6: ...ely connected to the item under test before pressing the Test button The DLRO200 checks voltage and current loop continuity as in Normal Mode and if adequate passes a continuous d c current repeating...

Страница 7: ...sing the two axis cursor control and the Enter key These are Retrieve Passbands Set Clock Delete Data and Storage Retrieve allows the recall of results which have been stored within the DLRO200 s inte...

Страница 8: ...ng may continue without Passbands The passband option allows you to set upper and lower limits between which the test result must lie if it is to be assigned a Pass Readings that fall above the upper...

Страница 9: ...sed Initially the Potential circuit is checked If there is poor continuity in this circuit a message will appear on the display stating POTENTIAL FAIL Press Enter to clear this message and correct the...

Страница 10: ...u do not wish to add notes press the Test button and a new test will be started or the Enter button to return to the Main Screen TEST RESULT STORAGE Each test is numbered sequentially dated and timed...

Страница 11: ...r these leads should be adequately insulated and sufficiently strong mechanically to support their own weight if long leads are being used The size of the current leads will limit the maximum resistan...

Страница 12: ...ent Ranges Full Scale Resolution Volts full scale Max Test Current 999 9 0 1 0 2 V 200 A 9 999 m 1 2 V 200 A 99 99 m 10 4 V 40 A 999 9 m 100 4 V 4 A Subject to supply voltage conditions see below Accu...

Страница 13: ...M Megger is a registered trademark DLRO200_UG_en_V04 0810 www actoolsupply com www actoolsupply com...

Отзывы: