background image

CONTINUITY 
Measuring Range: 

0,01Ω to 99,9Ω

(0 to 10Ω on analogue scale)

EN61557 Operating range: 0,10Ω to 99,9Ω

Accuracy: ±2% 

±2 

digits

Open circuit voltage: 

5V ±1V

Test current: 

210mA ±10mA (0-2Ω)

Zero offset at probe tips:  0,10Ω typical

Lead resistance zeroing:

Up to 9,99Ω

Noise rejection:

1V rms 50/60Hz

Buzzer: 

Operates at less than 5Ω (approx).

RESISTANCE

Measuring Range: 

0,01kΩ to 9,99MΩ

(0 to 100MΩ on analogue scale)

Accuracy: 

±3%  ±2 digits

Open circuit voltage: 

5V ±1V

Short circuit current: 

25µA ±5µA

VOLTAGE

Measuring Range:

±1V to ±500V 

(0 to 1000V on analogue scale)

Accuracy: 0 to 500V d.c. ±2% ±3 digit

0 to 500V a.c (50/60Hz) ±2% ±3 digits

0-500V 400Hz a.c. ± 5% ±3 digits

MILLIVOLTS

Measuring Range:

±0,1mV to ±1999mV 

(0 to 1000mV on analogue scale)

Accuracy: 0,1mV to 10mV d.c. or a.c. (50/60Hz) ±2% ±5 digits

10mV to 1999mV d.c. or a.c. (50/60 Hz) ±2% ±3 digits

0,1mV to 10mV a.c. (16-460 Hz) ±5% ±7 digits

10mV to 1999mV a.c. (16-460 Hz)  ±5% ±5 digits

d.c. milliVolts zeroing:

Up to 9,9mV

Input resistance:

>3MΩ

CAPACITANCE 
Measuring Range:

0,1nF to 9,99uF

Accuracy:

±3% ±2digits ± 0,2nF

uF zeroing:

up to 10nF

Basic and service errors for Insulation and Resistance ranges

The basic error is the maximum inaccuracy of the instrument

under ideal conditions, whereas the service error is the

maximum inaccuracy taking into effect of battery voltage,

temperature, interference, and system voltage and frequency,

where applicable. After determining the service error, we can

then calculate the measurement range. This is the range of

measurement over which the error in service is less than 30%

of the reading. Digital instruments are affected by the number

of digits error – for example a value 0,10Ω measured with the

continuity range may give a display in the range 0,07Ω to

0,13Ω which is a maximum error of 30%. Therefore the

measurement range measuring low resistance is 0,10Ω to

99,9Ω. When checking that a measurement does not exceed a

limit, the service error needs to be taken into account and

Содержание BMM80 Series

Страница 1: ...ve no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high energ...

Страница 2: ...he instrument enables the optional Megger SP6F Switched Probe to be used for two handed probe operation The 250V 500V and 1000V ranges can be used to test electrical installations in compliance with B...

Страница 3: ...strument is switched to the OFF position when not in use Backlight The backlight is activated by pressing the key The backlight will remain illuminated for approximately one minute before automaticall...

Страница 4: ...Testing and value assessment can be found in Megger Limited publications listed in the Accessories page Automatic Discharge When the TEST button is released after an insulation test or re pressed if...

Страница 5: ...ossible sources of error Measurements and results can be effected by the following The impedance of operating circuits connected in parallel Impedance such as inductors that vary during the measuremen...

Страница 6: ...F The measured capacitance is indicated on the display 1 Set the selector switch to uF 2 Connect the test leads 3 After a short settling time the reading will be displayed automatically Zeroing of uF...

Страница 7: ...over Remove the cells if the instrument is not going to be used for an extended period of time Fuse Checking and Replacement To check the instrument fuse switch to an insulation range and press the TE...

Страница 8: ...1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M CAPACITANCE Measuring Range 0 1nF to 9 99uF Accuracy 3 2digits 0 2nF uF zeroing up to 10nF Basic and service erro...

Страница 9: ...00 4 76 0 50 0 44 10 00 9 56 0 60 0 54 20 00 18 8 0 70 0 63 30 00 28 4 0 80 0 73 40 00 38 0 0 90 0 82 50 00 47 6 1 00 0 92 100 00 92 0 SAFETY The instruments meet the requirements for double insulatio...

Страница 10: ...erating humidity 90 RH at 40 C max Storage temperature range 25 to 65 C Calibration Temperature 20 C Maximum altitude 2000m Dust and water protection IP54 Temperature coefficient 0 1 per C WEIGHT 742g...

Страница 11: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 12: ...l U S A Fax 1 214 331 7399 Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES Tel 33 0 1 30 16 08 90 FRANCE Fax 33 0 1 34 61 23 77 This instrument is manufactured in the United Kingdom Th...

Отзывы: