‘locked on’ by pressing the

key while the TEST

button is held down. The

warning

will appear on

the display and both

buttons may be released

whilst the test continues.

The next press of the test

button will terminate the

test.
Note:  There is a safety
delay of a few seconds on
the first operation of
‘1000V’ range, each time
the range is selected.
Polarisation Index Testing
Polarisation Index (PI) is the term applied to the Dielectric

Absorption Ratio when resistance values are measured after 1

minute and again after 10 minutes. Polarisation Index is then the

resistance value after 10 minutes divided by the resistance value

after 1 minute. The test can be run at any voltage. More detailed

information on PI Testing and value assessment can be found in

Megger Limited publications listed in the Accessories page.
Automatic Discharge
When the TEST button is released after an insulation test (or

re-pressed if ltb feature is enabled), a 200kΩ load is

automatically switched across the terminals to discharge the

item under test. Any voltage present will be indicated on the

display so that the discharge can be monitored. 

TYPICAL TERMINAL VOLTAGE CHARACTERISTICS

Continuity Testing (Ω) (See fig.2)
The continuity tests are activated when the probes make

contact of less than a few kΩ. The tests apply a constant

current and measure the resulting volt drop across the circuit

under test. The test operates without the need to press the

TEST button. When the test leads are removed the reading will

hold for a few seconds and then reset. This range is not

suitable for diode testing since the automatic contact detector

FIG.1

Содержание BMM80 Series

Страница 1: ...ve no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high energ...

Страница 2: ...he instrument enables the optional Megger SP6F Switched Probe to be used for two handed probe operation The 250V 500V and 1000V ranges can be used to test electrical installations in compliance with B...

Страница 3: ...strument is switched to the OFF position when not in use Backlight The backlight is activated by pressing the key The backlight will remain illuminated for approximately one minute before automaticall...

Страница 4: ...Testing and value assessment can be found in Megger Limited publications listed in the Accessories page Automatic Discharge When the TEST button is released after an insulation test or re pressed if...

Страница 5: ...ossible sources of error Measurements and results can be effected by the following The impedance of operating circuits connected in parallel Impedance such as inductors that vary during the measuremen...

Страница 6: ...F The measured capacitance is indicated on the display 1 Set the selector switch to uF 2 Connect the test leads 3 After a short settling time the reading will be displayed automatically Zeroing of uF...

Страница 7: ...over Remove the cells if the instrument is not going to be used for an extended period of time Fuse Checking and Replacement To check the instrument fuse switch to an insulation range and press the TE...

Страница 8: ...1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M CAPACITANCE Measuring Range 0 1nF to 9 99uF Accuracy 3 2digits 0 2nF uF zeroing up to 10nF Basic and service erro...

Страница 9: ...00 4 76 0 50 0 44 10 00 9 56 0 60 0 54 20 00 18 8 0 70 0 63 30 00 28 4 0 80 0 73 40 00 38 0 0 90 0 82 50 00 47 6 1 00 0 92 100 00 92 0 SAFETY The instruments meet the requirements for double insulatio...

Страница 10: ...erating humidity 90 RH at 40 C max Storage temperature range 25 to 65 C Calibration Temperature 20 C Maximum altitude 2000m Dust and water protection IP54 Temperature coefficient 0 1 per C WEIGHT 742g...

Страница 11: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 12: ...l U S A Fax 1 214 331 7399 Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES Tel 33 0 1 30 16 08 90 FRANCE Fax 33 0 1 34 61 23 77 This instrument is manufactured in the United Kingdom Th...

Отзывы: