background image

FUSE 

500mA (F) 500V, 32x 6mm Ceramic HBC 10kA minimum.

E.M.C.

The instruments meet EN 61326-1.

POWER SUPPLY

Battery Type:

6x1,5V Alkaline cells IEC LR6 type

or 1.2V NiCd re-chargeable cells.

Battery Life (typical):

2100 5-sec 1kV insulation tests

3200 5-sec 500V insulation tests

4000 5-sec 250V insulation tests

2700 5-sec continuity tests

4700 5-sec kΩ tests

ENVIRONMENTAL CONDITIONS

Operating range: 

-5 to +40°C

Operating humidity: 

90% RH at 40°C max.

Storage temperature range:

-25 to +65°C

Calibration Temperature:

+20°C

Maximum altitude:

2000m

Dust and water protection:

IP54

Temperature coefficient: 

<0,1% per °C 

WEIGHT

742g

DIMENSIONS

110mm x 220mm x 45mm

CLEANING

Wipe with a clean cloth

damped with soapy water or

Isopropyl Alcohol (IPA)

ACCESSORIES

Supplied: Part 

Number

Test lead set

6220-437

Test and Carry case

6420-123

Optional:

Part Number

Fused lead set, FPK8 

6111-218

Switch Test Probe SP6F 

6220-836

Test Record Cards (Pack of 20) 

6111-216

Publications:

‘A Stitch in Time’

AVTM21-P8B

‘Testing Electrical Installations’

6231-605

Содержание BMM80 Series

Страница 1: ...ve no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high energ...

Страница 2: ...he instrument enables the optional Megger SP6F Switched Probe to be used for two handed probe operation The 250V 500V and 1000V ranges can be used to test electrical installations in compliance with B...

Страница 3: ...strument is switched to the OFF position when not in use Backlight The backlight is activated by pressing the key The backlight will remain illuminated for approximately one minute before automaticall...

Страница 4: ...Testing and value assessment can be found in Megger Limited publications listed in the Accessories page Automatic Discharge When the TEST button is released after an insulation test or re pressed if...

Страница 5: ...ossible sources of error Measurements and results can be effected by the following The impedance of operating circuits connected in parallel Impedance such as inductors that vary during the measuremen...

Страница 6: ...F The measured capacitance is indicated on the display 1 Set the selector switch to uF 2 Connect the test leads 3 After a short settling time the reading will be displayed automatically Zeroing of uF...

Страница 7: ...over Remove the cells if the instrument is not going to be used for an extended period of time Fuse Checking and Replacement To check the instrument fuse switch to an insulation range and press the TE...

Страница 8: ...1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M CAPACITANCE Measuring Range 0 1nF to 9 99uF Accuracy 3 2digits 0 2nF uF zeroing up to 10nF Basic and service erro...

Страница 9: ...00 4 76 0 50 0 44 10 00 9 56 0 60 0 54 20 00 18 8 0 70 0 63 30 00 28 4 0 80 0 73 40 00 38 0 0 90 0 82 50 00 47 6 1 00 0 92 100 00 92 0 SAFETY The instruments meet the requirements for double insulatio...

Страница 10: ...erating humidity 90 RH at 40 C max Storage temperature range 25 to 65 C Calibration Temperature 20 C Maximum altitude 2000m Dust and water protection IP54 Temperature coefficient 0 1 per C WEIGHT 742g...

Страница 11: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 12: ...l U S A Fax 1 214 331 7399 Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES Tel 33 0 1 30 16 08 90 FRANCE Fax 33 0 1 34 61 23 77 This instrument is manufactured in the United Kingdom Th...

Отзывы: