4 FUNCTIONS AND SET UP
4
F
unctions and set up
4.1 General
Internal software version
When turning the function selector from OFF to any
position the display will shortly show a number e.g.
008 and then during five seconds the internal soft-
ware version e.g. "01C" (revision R01C).
OFF
The instrument should be in the
OFF
position
when
not in use.
Note
The MOM2 will go to stand by mode after
about 10 minutes of inactivity. To wake up,
press Stand-by / Wake up (CLR LOG) button.
4.2 Test positions
I > I min
Minimum current is set in position
SET / I min
.
Generation / measurement time: 0.1 s, 0.6 s or 3 s.
I = I max
The current is only limited by the total impedance of
the circuit.
Generation / measurement time: 0.1 s, 0.6 s or 3 s.
USER / 1, 2, 3
User defined settings as generation time, I min, pass/
fail and log.
Note
The user defined settings can only be set via
MOM2 Win from a PC