14

Immunity 

test

IEC 60601 

test level

Compliance 

level

Electromagnetic 

environment - guidance

Guidance and MANUFACTURER’S declaration – electromagnetic 

immunity – for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity

The model ESM201 is intended for use in the electromagnetic environment 

specified below. The customer or the user of the model ESM201 should assure that 

it is used in such an environment.

Radiated 

RF

IEC 

61000-4-3

Portable and mobile RF communications 

equipment should be used no closer to 

any part of the model ESM201, including 

cables, than the recommended separation 

distance calculated from the equation 

applicable to the frequency of the 

transmitter.

Recommended separation distance 

where P is the maximum output power 

rating of the transmitter in watts (W) 

according to the transmitter manufacturer 

and d is the recommended separation 

distance in metres (m).b

Field strengths from fixed RF transmitters, 

as determined by an electromagnetic site 

survey, a should be less than the 

compliance level in each frequency range.b

Interference may occur in the 

vicinity of equipment marked 

with the following symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by 

absorption and reflection from structures, objects and people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 

environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the model ESM201 is used 

exceeds the applicable RF compliance level above, the model ESM201 should be observed to 

verify normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as re-orienting or relocating the model ESM201.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

3 V/m

80 MHz to 2.5 

GHz

3 V/m

15

Recommended separation distances between portable and mobile RF 

communications equipment and the EQUIPMENT or SYSTEM - for 

EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Separation distance according to frequency of 

transmitter

m

Recommended separation distances between

portable and mobile RF communications equipment and the model ESM201

Rated maximum 

output of 

transmitter

W

150 kHz to 80 

MHz 

80 MHz to 800 

MHz

80 MHz to 800 

MHz

The model ESM201 is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the model 

ESM201 can help prevent electromagnetic interference by maintaining a minimum 

distance between portable and mobile RF communications equipment 

(transmitters) and the model ESM201 as recommended below, according to the 

maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above the recommended 

separation distance d in metres (m) can be estimated using the equation applicable to the 

frequency of the transmitter, where P is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range 

applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.

0.01

0.1

1

10

100

0.12

0.38

1.2

3.8

12

0.12

0.38

1.2

3.8

12

0.23

0.73

2.3

7.3

23

P

V

d

]

5

.

3

[

1

=

P

E

d

]

5

.

3

[

1

=

P

E

d

]

7

[

1

=

Содержание ESM201

Страница 1: ...is the standard Shenzhen Med link Electronics Tech Co Ltd ShanghaiInternationalHoldingCorp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Tel 49 40 2513175 Fax 49 40 255726 R E V A 0 4th and 5th Fl...

Страница 2: ......

Страница 3: ...e for neonatal infant and the people who can t express his her thought This device is used to non invasive measure human body s blood pressure Do not use it for other purpose Do not use this device cl...

Страница 4: ...CD full display Cuff connector Cuff Air hose 2 ON OFF button MEMORY button SET button USER button DOWN button UART port LCD display Cuff socket Battery compartment cover 3 1 Install or Replace Batteri...

Страница 5: ...Measurement To ensure the measurement accuracy please follows the below suggestions Do not eat drink smoke or exercise before measurement Take the measurement after urination and defecation Sit down...

Страница 6: ...er measurement is finished the cuff will deflation automatically then the systolic diastolic blood pressure pulse rate and blood pressure unit will be displayed and stored If the voice guided function...

Страница 7: ...r and get into the food chain damaging your health and well being 7 Maintenance and Storage 9 The below table explains the error code that may display on the LCD The below table explains the error cod...

Страница 8: ...IEC EN 60601 1 EMC IEC EN 60601 1 2 performance EN1060 1 3 IEC EN 80601 2 30 User manual Warranty card Gift box 80 105kPa 2x90 sets Temperature 20 C 55 C Humidity 95 RH non condensing The following s...

Страница 9: ...blishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Emissions test Compliance El...

Страница 10: ...ield strength in the location in which the model ESM201 is used exceeds the applicable RF compliance level above the model ESM201 should be observed to verify normal operation If abnormal performance...

Страница 11: ...ure exerted by circulating blood upon the walls of blood vessels and is one of the principal vital signs Two pressures are measured for a blood pressure reading Systolic blood pressure is a measure of...

Страница 12: ......

Отзывы: