The Pump performs a series of safety checks immediately after the two
parts are connected. The safety checks take about twenty seconds. The
indicator light flashes in the order of blue, green, yellow, and red, and the
Pump beeps four times.
Warning:
If the Pump fails to beep or the indicator light fails to flash, call
customer support. It may put your health at risk if you continue to use the Pump.
Warning:
If a single fault condition occurs, the indicator light flashes red, and
the Pump cannot proceed to the next step, call customer service.
4.1.4 Fill the New Reservoir Patch
Warning:
If your insulin is stored in the refrigerator, wait until the insulin
reaches room temperature before you fill the reservoir. Using cold insulin could
produce air bubbles in the reservoir. While filling the reservoir, be sure to
remove air bubbles.
Warning:
When you are filling the Reservoir Patch, make sure that it is at least
30 cm (12 inches) from any magnetic objects, such as magnets, mobile phones,
and other Reservoir Patches. The Patch Pump will detect the volume of insulin
in the reservoir once it is filled, and if the Patch Pump is in a magnetic field, the
volume detected can be inaccurate.
Take the following steps to fill a new Patch:
1.
Clean the top of the insulin vial with an alcohol wipe.
2.
Remove the protective cap from the needle. Keep the cap.
B ureau Veritas