112
A specified number of units that remain in the Reservoir Patch
A specified maximum amount of time that remains before the
Reservoir Patch will be empty
Note:
You can set the amount of insulin between 5 U and 50 U with an
increment of 1 U. You can set the time between 2h and 24h with an increment
of 30 min.
Note:
If a bolus is being delivered when a “LOW RESERVOIR
”
Alert occurs, your
remaining insulin may be less than the value on the alert screen.
3. Max Delivery
This program is designed to instruct the Pump to automatically suspend insulin
delivery and give an alarm if you may have delivered excessive insulin in the past
hour or within one day. There are two types of delivery limit that you can set,
hour max and daily max. The factory setting for hour max is 25 U, and for daily
max is 80 U. You may choose to program this feature into your PDM based on
the amount of units you usually deliver in 1 hour and within 1 day. Discuss what
settings are best with your healthcare provider.
Note:
You can set the Daily Max between 20 and 180 U and the Hourly Max
between 10 and 40U.
Note:
If the Reservoir Patch is changed, the hourly insulin amount will start
from 0.
Note:
Insulin delivery will be automatically resumed at 0:00 am on the next
day if suspension occurs after daily max has been exceeded.