Sterilizing
The adapters, manual retraction tool, and the reusable
insertion guide are provided non-sterile. They may be
sterilized by steam autoclave.
WARNING
Do not use hydrogen peroxide gas plasma technology
(such as STERRAD
™*
systems), ethylene oxide, or gamma
sterilization. The adapters and accessories are approved for
steam autoclave sterilization.
The following information is the sterilization method
recommended and qualified for the linear adapter. Do not
expose the device to temperatures in excess of 279 F
(137 C), as this may shorten device service life and/or lead to
device failure.
∙
Place the adapter on its side during sterilization.
∙
Allow a 20-minute cool down period at room temperature
post sterilization. Do not leave the instrument in the
autoclave for cool down; remove it immediately after the
sterilization cycle is complete.
∙
Do not use flash steam sterilization. Use of flash steam
sterilization will damage the device, and may lead to
malfunction.
Steam Autoclave Sterilization
132 C pre-vacuum (Hi Vac) steam cycle
∙
Exposure temperature: 270 F (132 C)
∙
Exposure time: 4 minutes
∙
Vacuum dry time: 20-40 minutes
134 C pre-vacuum (Hi Vac) steam cycle
∙
Exposure temperature: 273 F (134 C)
∙
Exposure time: 3 minutes
∙
Vacuum dry time: 20-40 minutes
Outside USA (OUS)
Minimum recommended WHO
†
Cycle
Pre-Vacuum Steam Cycle
Cycle
Cycle
Cycle
Exposure Temperature (C)
132
134
134
134
Exposure Time (minutes)
4
3
5
18
Vacuum Dry Time (minutes) 20-40
20-40
20-40 20-40
†World Health Organization
NOTE
When sterilizing multiple instruments in one autoclave
cycle, ensure the sterilizer manufacturer’s stated maximum
load is not exceeded.
Storage
Store at room temperature.
Transport humidity
limitation
90%
Transport atmospheric
pressure
54.9 kPa
101.3 kPa
20.2°F
-29°C
140°F
60°C
Transport temperature
limitations
Disposal
Discard or recycle as per local, state, and governmental
regulations.
STERILIZATION
INSTRUCTIONS