Manual cleaning
1. Remove the adapter from the stapler according to the
instructions for use provided with the Signia
™
stapling
system’s user manual.
2. Wipe down the adapter with a separate lint-free cloth
soaked in a 90-104 F (32-40 C) pH-neutral detergent
solution diluted per the manufacturer’s specifications.
3. Under 90-104 F (32-40 C) running water, scrub all
reachable exterior surfaces with a general instrument
soft nylon bristle brush for 1-2 minutes to remove
surgical debris. Hold the distal end of the adapter under
running water and flush for 1-2 minutes. Do not insert
the brush into the distal end shaft. While brushing, pay
particular attention to the grooves on the pin. Also pay
particular attention to the grooves and sides of the reload
unload button.
4a. For enzymatic detergents (Validated with Steris
Prolystica
™*
2X Concentrate): Hold the adapter at a slight
angle to allow the detergent solution to flow into the shaft
from the bottom and air bubbles to release from the top.
Once all air bubbles are released, immerse the adapter in
a 90-104 F (32-40 C) enzymatic bath diluted as specified
by the manufacturer’s Instructions for Use and soak for
10-20 minutes.
4b. For alkaline detergents (non-enzymatic pH greater
than 9.5 but less than 10.8) (Validated with neodisher
MediClean forte
™*
): Hold the adapter at a slight angle to
allow the detergent solution to flow into the shaft from
the bottom and air bubbles to release from the top.
Once all air bubbles are released, immerse the adapter
in a 90-104 F (32-40 C) alkaline bath diluted as specified
by the manufacturer’s Instructions for Use and soak for
10-20 minutes.
5. Manually agitate the adapter in the bath for 1-2 minutes.
6. Invert the adapter until all fluid has completely drained.
7. In a clean 90-104 F (32-40 C) water bath, hold the adapter
at a slight angle to allow the water to flow into the shaft
from the bottom and air bubbles to release from the top.
Once all air bubbles are released, immerse the adapter
and soak for 2-3 minutes.
8. Manually agitate the adapter in the bath for 1-2 minutes.
9. Invert the adapter until all fluid has completely drained.
10. Repeat step 3.
11. Rinse under 90-104 F (32-40 C) running tap water for
1-2 minutes.
12. Perform a final rinse under purified water for 1-2 minutes.
13. Dry with a clean, soft, lint-free cloth.
14. Inspect the adapter. If not visibly clean, repeat the
above steps.
PRECAUTION
Do not use instrument lubricant on the power handle or the
adapter.
Automatic cleaning
1. Remove the adapter from the stapler according to the
instructions for use provided with the Signia
™
stapling
system’s user manual.
2. Wipe down the adapter with a lint-free cloth soaked in 90-
104 F (32-40 C) pH-neutral detergent solution diluted per
manufacturer’s instructions.
3. Under 90-104 F (32-40 C) running water, scrub all
reachable surfaces with a general instrument soft nylon
bristle brush for 1-2 minutes to remove surgical debris.
Hold the distal end of the adapter under running water
and flush for 1-2 minutes. Do not insert the brush into the
distal end shaft. While brushing, pay particular attention to
the grooves on the pin. Also pay particular attention to the
grooves and sides of the reload unload button.
4. Rinse under 90-104 F (32-40 C) running tap water for
1-2 minutes.
5. Perform the automatic cleaning cycle, following the
recommendations in the table on the next page.
NOTE
The Signia
™
adapter should be placed in such a manner to
avoid contact with other devices to prevent damage from
occuring in result of movement during the wash cycle. It is
recommended that the adapter be placed with reload unload
button facing downward to assist drainage.
Treatment
Time (min:sec) Temperature
Chemicals
Pre-Wash
00:45
Cold tap water
N/A
Wash
04:00
Hot tap water
Enzymatic
1
or Alkaline
2
Detergent
3
Rinse
00:15
Hot tap water
N/A
Wash
03:00
Hot tap water
Enzymatic
1
or Alkaline
2
Detergent
3
Rinse
00:15
Hot tap water
N/A
Thermal Rinse 05:00
Hot purified
water heated to
203 F (95 C)
N/A
Dry
06:00
High setting
203 F (95 C)
N/A
1. Validated with Steris Prolystica
™*
2x concentrate
2. Validated with neodisher MediClean forte
™*
3. Dilute detergents per the manufacturer’s specifications
6. Dry with a clean, soft, lint-free cloth.
7. Inspect the adapter. If not visibly clean, repeat the above
steps.
PRECAUTION
Do not use instrument lubricant on the power handle or the
adapter.